The Recall Desk

Hazard

Brake Interference recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
0
Most recent
2021-12-13

Of the 1 brake interference recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all brake interference recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • HighNHTSA·21V966000·2021-12-13

    Autocar Xpeditor GPS Module Mounting Defect May Interfere With Brakes

    Autocar is recalling certain 2021-2022 Xpeditor vehicles because the GPS module may be improperly mounted and could come loose, potentially interfering with brake and throttle pedals.

    Product
    AUTOCAR — AUTOCAR XPEDITOR
    Category
    Vehicle
    Distribution
    Distributed nationwide