The Recall Desk

Hazard

Analytical Error recalls

8 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
8
Critical
0
Severe
0
Most recent
2025-07-02

Of the 8 analytical error recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all analytical error recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–8 of 8

  • ModerateFDA (Devices)·Z-2027-2025·2025-07-02

    Atellica CH Diluent Contaminated with Sodium Hypochlorite

    Siemens Healthcare Diagnostics is recalling Atellica CH Diluent (SMN 11099300, Lot 0000195412) due to contamination with sodium hypochlorite, which may affect laboratory test results.

    Product
    Product: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1993-2025·2025-06-25

    STA-Owren Koller Buffer Solution Recall Due to Clotting Time Concerns

    Diagnostica Stago is recalling STA-Owren Koller buffer solution (Lot 270902) because it may produce shorter clotting times in certain tests, potentially leading to a 1% overestimate of clotting results.

    Product
    STA - Owren Koller. Model/Catalog Number: 00360 STA - Owren-Koller is a buffer solution intended for use as a diluent for reagents and human plasmas in coagulation tests performed with analyzers of the STA-R, STA Compact, STA Satellite and ST art families.
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-1549-2024·2024-04-24

    Siemens Atellica Lab Analyzers Risk Falsely Elevated Cholesterol Results

    Siemens Atellica CH and CI analyzers may produce falsely elevated cholesterol and lipid test results (2–16% high) after iron tests. This affects clinical calibration, quality control checks, and patient test results.

    Product
    Atellica CH Iron3 - IVD use in the quantitative measurement of iron in human serum and plasma (lithium heparin and sodium heparin) using the Atellica¿ CI Analyzer and Atellica¿ CH Analyzer Siemens Material Number (SMN): 11537211
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1572-2023·2023-05-17

    APLS IgM Diagnostic Reagent Packs Recalled Due to Incorrect Conjugate

    Bio-Rad Laboratories is recalling 932 BioPlex 2200 APLS IgM Reagent Packs (Lot 301538) nationwide because they were packaged with incorrect conjugate. This could result in false-positive and false-negative test results.

    Product
    BioPlex 2200 REF 665-2050, APLS IgM Pack, APLS IgM Reagent Pack
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1305-2023·2023-04-05

    ACL ELITE PRO-Analyzer: HemosIL Liquid Anti-Xa test definition removed

    Instrumentation Laboratory is removing the HemosIL Liquid Anti-Xa test definition from ACL ELITE/Elite Pro analyzers used in clinical hemostasis laboratories. No illnesses or injuries have been reported.

    Product
    ACL ELITE PRO-Analyzer designed specifically for clinical use in the hemostasis laboratory, for coagulation and/or fibrinolysis testing. Part Number: 0001000311
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2184-2021·2021-08-11

    Abbott Alinity c Reagent Kit Recall Due to Potential Test Result Errors

    Abbott Laboratories is recalling its Alinity c Activated Alanine Aminotransferase reagent kit due to potential analytical errors that could produce incorrect test results. The issue affects certain lot numbers distributed worldwide.

    Product
    REF: 08P1820, Abbott Alinity c Activated Alanine Aminotransferase reagent Kit, Alinity c Activated Alanine Aminotransferase Reagent Kit,, IVD, CE, UDI: (01) 00380740132569
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2966-2020·2020-09-23

    Randox Ammonia Reagent Recall Due to Positive Bias in Patient Samples

    Randox Laboratories is recalling 143 kits of the AM1054 Ammonia reagent because it shows a positive bias of up to 140 µmol/l on patient samples.

    Product
    AM1054 Randox Ammonia (NH3), Enzymatic UV Method For the quantitative in vitro determination of Ammonia in plasma
    Category
    Medical Device
    Distribution
    3 states
  • ModerateFDA (Devices)·Z-1532-2014·2014-05-07

    Siemens Androstenedione Assay Recall Due to Over-Recovery

    Siemens recalls Androstenedione assays because they over-recover samples above 5.5 ng/mL, contradicting the stated 10 ng/mL limit.

    Product
    IMMULITE 2000/2000 XPi Systems Androstenedione (Test Code AND), REF L2KAO2, SMN 10381188, IVD, 200 test kit size --- Siemens Healthcare Diagnostics Products Ltd. Llanberis, Gwynedd, LL55 4EL, UK For in vitro diagnostic use with the IMMULITE/IMMULITE 1000 and 2000 Systems Ana
    Category
    Medical Device
    Distribution
    Distributed nationwide