Philips Tempus Pro Patient Monitor Software Error Impairs Critical Monitoring
Certain Philips Tempus Pro Patient Monitors may display an error message when laryngoscopes are disconnected, requiring a 60-100 second restart that prevents vital sign monitoring and airway visualization during intubation. This may lead to treatment delays or hypoxia.
- Product
- Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, and 00-1007, when Used With Tempus Pro Video Laryngoscope, REF 01-2044
- Category
- Medical Device
- Distribution
- 42 states