Daavlin Distributing Company recalled 3 products on 2026-07-22

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2026-07-22
Critical
0
Units
240

Why these products were recalled

Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • ModerateFDA (Devices)··2026-07-22

    Photera 600 Phototherapy Device UV Sensor Calibration Error

    Daavlin Distributing Company is recalling 26 Photera 600 phototherapy devices nationwide because the UV sensor gain calibration is incorrect, causing inaccurate ultraviolet output readings.

    Product
    Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2026-07-22

    Photera 600 3-D phototherapy device UV sensor calibration error recall

    Daavlin Distributing Company is recalling Photera 600 3-D phototherapy devices because the UV sensor gain calculation is incorrect, causing inaccurate ultraviolet output measurement.

    Product
    Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2026-07-22

    Photera 400 Phototherapy Device Recalled for Incorrect UV Sensor Calibration

    Daavlin Distributing Company is recalling the Photera 400 phototherapy device because the UV sensor calibration is calculating incorrectly, resulting in lower output levels than intended.

    Product
    Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders