The Recall Desk

FDA (Devices) recall action 99279

GE Medical Systems, LLC recalled 3 products on 2026-07-22

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2026-07-22
Critical
0
Units
66

Why these products were recalled

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • HighFDA (Devices)··2026-07-22

    GE Optima XR200/220/240amx Radiographic System Column Safety Mechanism

    GE Medical Systems is recalling certain Optima XR200/220/240amx mobile radiographic systems after column replacements because functional testing of the safety pawl may not have been performed, which could allow the X-ray arm to drop and injure patients.

    Product
    Optima XR220amx, Mobile Digital-Ready Radiographic System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2026-07-22

    Mobile Digital Radiographic System Arm Drop Risk During Patient Exams

    GE HealthCare is recalling certain Optima XR200/220/240amx Mobile Digital Radiographic Systems due to potential failure of the Column Safety mechanism, which could cause the X-ray arm to drop during patient exams. The issue stems from incomplete functional testing of the safety pawl after column replacement.

    Product
    Optima XR240amx, Mobile Digital Radiographic System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2026-07-22

    GE Medical Optima XR200/220/240amx Mobile Radiographic Systems Recall

    GE HealthCare recalls certain Optima XR200/220/240amx mobile radiographic systems where safety testing may have been incomplete after column replacement. A failed counterpoise cable could cause the X-ray arm to drop on patients.

    Product
    Optima XR200amx, Mobile Digital-Ready Radiographic System
    Category
    Distribution
    0 states