FDA (Devices) recall action 98992

Siemens Medical Solutions USA, Inc recalled 2 products on 2026-06-10

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2026-06-10
Critical
0
Units
—

Why these products were recalled

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

The agency’s own wording, quoted from the enforcement report.

1–2 of 2

  • HighFDA (Devices)··2026-06-10

    ARTIS Icono systems may display zero dose while dose is applied

    ARTIS Icono interventional fluoroscopic X-ray systems are being recalled because they may sporadically display a zero X-ray dose even though an actual dose is applied to the patient.

    Product
    ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional fluoroscopic x-ray system
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2026-06-10

    ARTIS pheno systems may display zero dose while dose is applied

    ARTIS pheno interventional fluoroscopic X-ray systems are being recalled because they may sporadically display a zero X-ray dose even though an actual dose is applied to the patient.

    Product
    ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system
    Category
    Distribution
    0 states