The Recall Desk

FDA (Devices) recall action 98642

Medline Industries, LP recalled 42 products on 2026-05-13

The FDA (Devices) publishes a recall like this one as 42 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 42 pages.

Products
42
Announced
2026-05-13
Critical
0
Units
86,259,492

Why these products were recalled

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

26–42 of 42

  • ModerateFDA (Devices)··2026-05-13

    Medline Convenience Kits surgical packs sterilization calibration recall

    Medline Industries is recalling three models of surgical convenience kits—Lithotomy Packs and Lap Basin—due to calibration issues with sterilization equipment that could affect the sterility assurance of the products.

    Product
    Medline Convenience Kits: 1) LITHOTOMY PACK, Model Number: DYNJ83185; 2) LITHOTOMY PACK, Model Number: DYNJ83185A; 3) LAP BASIN, Model Number: DYNJVB91109D
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2026-05-13

    Medline Neurosurgical Convenience Kits Sterilization Calibration Defect

    Medline is recalling 52 models of neurosurgical convenience kits due to calibration issues with sterilization equipment that may have compromised the sterility assurance level of the products.

    Product
    Medline Convenience Kits: 1) NEURO SHUNT- CODMAN, Model Number: CDS860018G; 2) NEURO SHUNT- CODMAN, Model Number: CDS860018I; 3) KIT NEUR FUS LUM POST CRANDALL, Model Number: DYKMBNDL138; 4) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142; 5) KIT NEURO FUS LUM…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2026-05-13

    Medline Tracheostomy Convenience Kits Sterilization Calibration Recall

    Medline Industries is recalling certain tracheostomy convenience kits and related products due to calibration issues with sterilization equipment that may affect the sterility assurance level of the devices.

    Product
    Medline Convenience Kits: 1) KIT STC TRACHEOSTOMY, Model Number:DYKMBNDL55; 2) KIT STC TRACHEOSTOMY, Model Number:DYKMBNDL55A; 3) TRACH TOTE, Model Number:DYNJ85691
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2026-05-13

    Medline Convenience Kits recalled due to sterilization calibration issues

    Medline Industries is recalling two models of Convenience Kits because calibration problems with sterilization and packaging equipment may have compromised the sterility of the products.

    Product
    Medline Convenience Kits: 1) DRAPE PACK, Model Number: DYNJ35363A; 2) SICK KIDS-GENERAL CUSTOM SPLIT, Model Number: SPTPCD10005
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2026-05-13

    Medline Surgical Drapes Recalled Due to Sterilization Equipment Calibration Issues

    Medline is recalling multiple models of surgical drapes due to calibration issues with sterilization and packaging equipment that may have compromised the sterility assurance level of the products. All affected units were distributed nationwide.

    Product
    Medline Surgical Drapes: 1) SHEET,DRAPE,40X58,STERILE, Model Number: DYNJP2410; 2) DRAPE,SHEET,ECONOMY,40X58",ST,20/CS, Model Number: DYNJP2411; 3) SHEET,DRAPE, 40X70,STERILE, Model Number: DYNJP2412; 4) DRAPE,MEDIUM,TAPE,40X70",ST,50/CS, Model Number: DYNJP2412T; 5)…
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2026-05-13

    Medline Convenience Kits recalled for potential sterility assurance defects

    Medline Industries is recalling six models of medical device convenience kits due to calibration issues with sterilization and packaging equipment that may have reduced sterility assurance levels. The recall affects 553 units distributed nationwide.

    Product
    Medline Convenience Kits: 1) PICC CATHETER INSERTION TRAY, Model Number: CVI4300; 2) OSC SHERMAN KNEE ARTHRO BASIC, Model Number: DYKM1977C; 3) KIT ACES INSERT PEG, Model Number: DYKMBNDL130; 4) KIT ACES INSERT PEG, Model Number: DYKMBNDL130A; 5) KIT ACES INSERT…
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2026-05-13

    Medline Convenience Kits recalled for potential sterilization issues

    Medline Industries is recalling multiple convenience kit models due to calibration issues with equipment used to sterilize and package the devices, which could affect sterility assurance levels.

    Product
    See RES for full list. Medline Convenience Kits: 1) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069G; 2) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069I; 3) HEAD & NECK CDS-LF, Model Number: CDS983782C; 4) BAPTIST FLOYD BRONCHOSCOPY, Model Number: DYKE1456D; 5)…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2026-05-13

    Medline Convenience Kits recalled due to sterilization calibration issues

    Medline Industries has recalled 6,676 units of multiple Medline Convenience Kits used in gynecological and other surgical procedures due to calibration issues with sterilization equipment that may affect sterility assurance.

    Product
    Medline Convenience Kits: 1) HYSTEROSCOPY, Model Number: CDS984086I; 2) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL91A; 3) VAGINAL HYSTERECTOMY PACK-LF, Model Number: DYNJ0373807J; 4) VAGINAL HYSTERECTOMY PACK-LF, Model Number: DYNJ0373807K; 5) D&C/HYSTEROSCOPY PACK…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2026-05-13

    Medline Cardiac Surgery Kits Recalled Due to Sterilization Equipment Calibration Issues

    Medline Industries is recalling cardiac surgery convenience kits because sterilization and packaging equipment calibration issues may have affected the sterility assurance level of the devices.

    Product
    See RES for complete list. Medline Convenience Kits: 1) OPEN HEART CDS, Model Number:CDS840015AQ; 2) OPEN HEART CDS, Model Number:CDS840015AQ; 3) OPEN HEART CDS, Model Number:CDS840023T; 4) OFF PUMP CABG CDS, Model Number:CDS840087AI; 5) OFF PUMP CABG CDS, Model…
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2026-05-13

    Medline Convenience Kits recalled due to sterilization equipment calibration issues

    Medline Industries is recalling four Convenience Kit models (892 units) nationwide because sterilization and packaging equipment calibration issues may have compromised the sterility assurance level of the devices.

    Product
    Medline Convenience Kits: 1) CHEST TUBE INSERTION TRAY, Model Number: DYNDA2941A; 2) CHEST/BREAST-LF, Model Number: DYNJ83950; 3) HERNIA REPAIR TOTE, Model Number: DYNJ908337A; 4) HERNIA REPAIR TOTE, Model Number: DYNJ908337C
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2026-05-13

    Medline Convenience Kits sterilization and packaging calibration issues recalled

    Medline Industries is recalling multiple Convenience Kit models used in cardiac, vascular, and surgical procedures due to calibration issues with sterilization and packaging equipment that could impact sterility assurance. Approximately 39,341 units distributed worldwide may be affected.

    Product
    Medline Convenience Kits: 1) ANGIO PROCEDURE PACK, Model Number:00-401824AB; 2) OR ARTERIOGRAM, Model Number:CDS985094F; 3) INTRAVASOR CATH KIT, Model Number:DYK1060193I; 4) KIT SURG ONC LAP DIAG/CV CATH, Model Number:DYKMBNDL153; 5) KIT ACES INSERT PERCUTANEOUS T…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2026-05-13

    Medline Urology Convenience Kits recalled due to sterilization calibration issues

    Medline Industries is recalling five models of urology convenience kits due to calibration issues in sterilization and packaging equipment that may compromise sterility assurance.

    Product
    Medline Convenience Kits: 1) KIT UROLOGY SLING, Model Number: DYKMBNDL106; 2) KIT UROLOGY SLING, Model Number: DYKMBNDL106A; 3) KIT UROLOGY PENILE REPAIR, Model Number: DYKMBNDL188; 4) KIT UROLOGY PERIPHERAL STIMULA, Model Number: DYKMBNDL189; 5) KIT UROLOGY PERIPHERAL…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2026-05-13

    Medline Burn Care Kits Recalled Due to Sterilization Calibration Issues

    Medline Industries is recalling three burn care convenience kits because calibration issues with sterilization and packaging equipment could compromise their sterility assurance level.

    Product
    Medline Convenience Kits: 1) BURN PACK, Model Number: DYNJ15668R; 2) BURN PACK-LF, Model Number: DYNJ42829B; 3) BURN CARE PACK-LF, Model Number: DYNJT3437
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2026-05-13

    Medline Surgical Convenience Kits recalled for sterilization calibration defect

    Medline Industries is recalling multiple surgical convenience kits because equipment used to sterilize and package the devices had calibration issues that may have affected sterility assurance. The devices were processed through validated sterilization cycles but the calibration problems could impact their sterility.

    Product
    See RES for complete list. Medline Convenience Kits: 1) SPINAL CDS, Model Number: CDS940087AG; 2) SPINAL CDS, Model Number: CDS940087AI; 3) TOTAL HIP CDS-LF, Model Number: CDS980832Q; 4) TOTAL HIP CDS-LF, Model Number: CDS980832R; 5) TOTAL HIP CDS-LF, Model Number: CDS980832S…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2026-05-13

    Medline Convenience Kits for C-Section and Obstetric Procedures

    Medline Industries is recalling 12,529 units of sterile surgical convenience kits used in cesarean section and obstetric procedures worldwide due to equipment calibration issues that may compromise sterilization assurance.

    Product
    Medline Convenience Kits: 1) C-SECTION PACK, Model Number: CDS820080N; 2) C-SECTION CDS, Model Number: CDS983523G; 3) C-SECTION, Model Number: CDSCCR611O; 4) C-SECTION, Model Number: CDSCCR611P; 5) LUD C-SECTION-LF, Model Number: DYNJ0101616I; 6) C-SECTION TRAY, Model…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2026-05-13

    Medline Surgical Packs recalled due to sterilization equipment calibration issues

    Medline Industries is recalling certain surgical packs because calibration issues with sterilization and packaging equipment may have affected sterility assurance. All recalled units were exposed to validated sterilization cycles but the calibration problems could impact sterility integrity.

    Product
    Medline Surgical Packs: 1) PACK,UNIVERSAL,ULTRAGARD,5/CS, Model Number: DYNJP1050UG; 2) PACK,SHOULDER,DRAPE,POUCH,10/CS, Model Number: DYNJP8401
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2026-05-13

    Medline Convenience Kits General Closure sterilization calibration issue

    Medline Industries is recalling Convenience Kits (Model DYKMBNDL200A) due to calibration issues in sterilization equipment that may have compromised sterility assurance. The kits were distributed nationwide.

    Product
    Medline Convenience Kits: 1) KIT GENERAL CLOSURE, Model Number: DYKMBNDL200A
    Category
    Distribution
    0 states