FDA (Devices) recall action 98635

Abbott Point Of Care Inc. recalled 3 products on 2026-05-06

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2026-05-06
Critical
0
Units
40,275

Why these products were recalled

Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, po…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • HighFDA (Devices)··2026-05-06

    i-STAT Blood Gas Analyzer Cartridges Recalled for Inaccurate Results

    Abbott is recalling certain i-STAT EG6+, EG7+, and G3+ blood gas cartridges that may report falsely high CO2 and falsely low pH results due to a manufacturing issue, potentially leading to unnecessary or harmful clinical interventions.

    Product
    i-STAT EG6+ cartridge; List Number: 03P77-25;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2026-05-06

    Abbott i-STAT EG7+ Blood Gas Cartridges Reporting Inaccurate Results

    Abbott Point Of Care is recalling approximately 7.6% of specific i-STAT EG7+ cartridge lots due to a manufacturing issue causing falsely high pCO2 and low pH readings. Inaccurate results could lead clinicians to make unnecessary or harmful medical interventions.

    Product
    i-STAT EG7+ cartridge; List Number: 03P76-25;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2026-05-06

    Abbott i-STAT Blood Gas Cartridges Reporting Inaccurate pH and PCO2 Results

    Abbott Point Of Care Inc. is recalling approximately 7.6% of i-STAT G3+, i-STAT EG6+, and i-STAT EG7+ blood gas cartridges due to a manufacturing issue causing falsely high PCO2 and low pH readings, which may lead to unnecessary or harmful clinical interventions.

    Product
    i-STAT G3+ cartridge; List Number: 03P78-26;
    Category
    Distribution
    0 states