FDA (Devices) recall action 98598

Medline Industries, LP recalled 19 products on 2026-04-15

The FDA (Devices) publishes a recall like this one as 19 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 19 pages.

Products
19
Announced
2026-04-15
Critical
0
Units
4,835,592

Why these products were recalled

Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–19 of 19

  • SevereFDA (Devices)··2026-04-15

    Medline robotic surgery packs recalled over syringe adaptor unwinding risk

    Medline is recalling Robotic Pack and Uro/Gen Robotic Pack convenience kits containing NAMIC rotating adaptor control syringes because the adaptor can unwind and disconnect during use.

    Product
    Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. ROBOTIC PACK, Medline SKU # DYNJ58475F; 2. URO/GEN ROBOTIC PACK, Medline SKU # DYNJ60803J.
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2026-04-15

    Medline TAVR procedure packs recalled over syringe adaptor disconnection risk

    Medline is recalling TAVR PACK convenience kits containing NAMIC rotating adaptor control syringes because the adaptor can unwind during use and disconnect from the manifold. 108 units are affected.

    Product
    Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: TAVR PACK, Medline SKU # DYNJ84472
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2026-04-15

    Medline CVOR angiography packs recalled over syringe adaptor unwinding risk

    Medline is recalling CVOR Angiography Pack-LF convenience kits containing NAMIC rotating adaptor control syringes because the adaptor can unwind and disconnect during use. 3,120 kits are affected.

    Product
    Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: CVOR ANGIOGRAPHY PACK-LF, Medline SKU DYNJ42367B
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2026-04-15

    Medline cath lab packs recalled over syringe rotating adaptor disconnection

    Medline is recalling cath lab and angiography procedure packs containing NAMIC rotating adaptor control syringes because the adaptor can unwind and disconnect. 17,902 kits are affected.

    Product
    Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. CATH LAB PACK-LF, Medline SKU 00-HCT055P; 2. INTERVENTIONAL CATH LAB PK-LF, Medline SKU DYNJ0376230Q; 3. ANGIOGRAPHY CATH…
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2026-04-15

    Medline minor plastic surgery packs recalled over syringe adaptor unwinding risk

    Medline is recalling Minor Plastic-LF convenience kits containing NAMIC rotating adaptor control syringes because the adaptor can unwind during use and disconnect from the manifold.

    Product
    Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: MINOR PLASTIC-LF, Medline Kit SKU # DYNJ86596
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2026-04-15

    NAMIC syringe recall reaches 2.6 million units over adaptor unwinding defect

    Medline is recalling a further tranche of NAMIC Angiographic Syringes because the rotating adaptor can unwind during use and disconnect from the manifold. 2,630,369 units are affected.

    Product
    NAMIC Angiographic Syringe, labeled as follows: 1. SYRINGE,20ML,CONTROL,-,PKG, Medline SKU 70075027; 2. SYRINGE,10ML,TR/FR,RA,W/RES, Medline SKU 70085007; 3. MTO,SYRINGE,10ML,TR/FG,RA,-,PKG, Medline SKU 70085107; 4. SYRINGE,12ML,TR/FR,RA,W/RES, Medline SKU 70087007…
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2026-04-15

    Medline delivery packs recalled over syringe rotating adaptor disconnection risk

    Medline is recalling Delivery Pack convenience kits containing NAMIC rotating adaptor control syringes because the adaptor can unwind during use and disconnect from the manifold.

    Product
    Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. DELIVERY PACK, Medline SKU # DYNJ42892F; 2. DELIVERY PACK, Medline SKU # DYNJ42892G.
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2026-04-15

    Medline head, neck and heart cath packs recalled over syringe adaptor defect

    Medline is recalling Head and Neck CDS-LF and Heart Cath Pack-LF convenience kits containing NAMIC rotating adaptor control syringes because the adaptor can unwind and disconnect.

    Product
    Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. HEAD AND NECK CDS-LF, Medline SKU # CDS780138K; 2. HEART CATH PACK-LF, Medline SKU # DYNJ36478B; 3. HEART CATH PACK-LF…
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2026-04-15

    Medline custom manifold kits recalled over syringe adaptor unwinding risk

    Medline is recalling custom manifold and cryo EP manifold kits containing NAMIC rotating adaptor control syringes because the adaptor can unwind and disconnect during use. 1,698 kits are affected.

    Product
    Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. CSTM MANIFOLD KIT - TITUS, Medline SKU VASC1061; 2. CSTM MANIFOLD KIT - FRYE, Medline SKU VASC1074; 3. CRYO EP MANIFOLD KIT…
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2026-04-15

    Medline special procedure trays recalled over syringe adaptor disconnection risk

    Medline is recalling Special Procedure Trays containing NAMIC rotating adaptor control syringes because the adaptor can unwind during use and disconnect from the manifold. 762 units are affected.

    Product
    Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: SPECIAL PROCEDURE TRAY, Medline SKU # DYNJ80513B
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2026-04-15

    Medline laparoscopic and minor procedure packs recalled over syringe adaptor defect

    Medline is recalling laparoscopic herniorrhaphy, lap and minor procedure packs containing NAMIC rotating adaptor control syringes because the adaptor can unwind and disconnect during use.

    Product
    Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. LAPAROSCOPIC HERNIORRHAPHY DSC, Medline SKU # DYNJ63021B; 2. LAP PACK, Medline SKU # DYNJ64297B; 3. MINOR PROCEDURE PACK…
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2026-04-15

    Medline angiography and neuro kits recalled over syringe adaptor disconnection

    Medline is recalling Angio Pack and Neuro Interventional Radiology convenience kits containing NAMIC rotating adaptor control syringes because the adaptor can unwind and disconnect during use.

    Product
    Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. ANGIO PACK, Medline SKU # DYNJ57032D; 2. NEURO INTERVENTIONAL RADIOLOGY, Medline SKU # DYNJ64026C; 3. NEURO INERVENTIONAL…
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2026-04-15

    Medline glove pack kits recalled over syringe adaptor disconnection risk

    Medline is recalling GLOVE PACK 7.0 convenience kits containing NAMIC rotating adaptor control syringes because the adaptor can unwind during use and disconnect from the manifold.

    Product
    Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: GLOVE PACK 7.0 (15PK), Medline SKU # DYNJ63673D
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2026-04-15

    NAMIC angiographic control syringes recalled over rotating adaptor unwinding

    Medline is recalling NAMIC Angiographic Control Syringes because the rotating adaptor can unwind during use, causing a loose connection or full disconnection from the manifold. 192,690 units are affected.

    Product
    NAMIC Angiographic Control Syringe, labeled as follows: 1. SYRINGE,7ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70083007; 2. SYRINGE,8ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70084007; 3. MTS,SYRINGE,8ML,TR/FR,RA,NO,RES,-,P/G, Medline SKU 70084017; 4.…
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2026-04-15

    Medline arthroscopy and extremity packs recalled over syringe adaptor unwinding

    Medline is recalling Shoulder Arthroscopy and Extremity Pack convenience kits containing NAMIC rotating adaptor control syringes because the adaptor can unwind and disconnect during use.

    Product
    Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. SHOULDER ARTHROSCOPY PACK, Medline SKU # DYNJ30880M; 2. EXTREMITY PACK, Medline SKU # DYNJ64295A.
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2026-04-15

    NAMIC angiographic syringes recalled in the largest rotating adaptor action

    Medline is recalling NAMIC Angiographic Syringes across multiple sizes because the rotating adaptor can unwind during use and disconnect from the manifold. 1,752,096 units are affected.

    Product
    NAMIC Angiographic Syringe, labeled as follows: 1. SYRINGE,20ML,CONTROL,-,PKG, Medline SKU 70075027; 2. SYRINGE,10ML,TR/FR,RA,W/RES, Medline SKU 70085007; 3. MTO,SYRINGE,10ML,TR/FG,RA,-,PKG, Medline SKU 70085107; 4. SYRINGE,12ML,TR/FR,RA,W/RES, Medline SKU 70087007; 5.…
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2026-04-15

    Additional NAMIC angiographic syringe lots recalled over adaptor unwinding

    Medline is recalling a further group of NAMIC Angiographic Syringe lots because the rotating adaptor can unwind during use and disconnect from the manifold. 149,439 units are affected.

    Product
    NAMIC Angiographic Syringe, labeled as follows: 1. SYRINGE,20ML,CONTROL,-,PKG, Medline SKU 70075027; 2. SYRINGE,10ML,TR/FR,RA,W/RES, Medline SKU 70085007; 3. MTO,SYRINGE,10ML,TR/FG,RA,-,PKG, Medline SKU 70085107; 4. SYRINGE,12ML,TR/FR,RA,W/RES, Medline SKU 70087007; 5.…
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2026-04-15

    Medline non-vascular and vein packs recalled over syringe adaptor defect

    Medline is recalling Non Vascular Pack and Vein convenience kits containing NAMIC rotating adaptor control syringes because the adaptor can unwind and disconnect during use.

    Product
    Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. NON VASCULAR PACK, Medline SKU # DYNJ68037C; 2. VEIN, Medline SKU # DYNJ89845; 3. VEIN, Medline SKU # DYNJ89845A.
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2026-04-15

    Medline cardiac catheterization kits recalled over syringe adaptor unwinding

    Medline is recalling right and left heart catheterization and angiography kits containing NAMIC rotating adaptor control syringes because the adaptor can unwind and disconnect. 79,843 units are affected.

    Product
    Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. MTS RH LFT HRT KIT, Medline SKU # 60010508; 2. MTS 3V W/GUIDE, Medline SKU # 60100701; 3. ANGIOGRAPHY KIT, Medline SKU #…
    Category
    Distribution
    2 states