The Recall Desk

FDA (Devices) recall action 98565

Straumann USA LLC recalled 2 products on 2026-04-01

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2026-04-01
Critical
0
Units
20

Why these products were recalled

A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2026-04-01

    Straumann custom abutments made from wrong bar material after mix-up

    Straumann is recalling Custom Abutment Ti for Straumann RC because a material mix-up meant titanium abutments were manufactured from TAN bar material instead of Titanium Grade 4.

    Product
    Custom Abutment Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 027.4620;
    Category
    Distribution
    10 states
  • HighFDA (Devices)··2026-04-01

    Straumann angled custom abutments made from wrong bar material

    Straumann is recalling Custom Abutment AS Ti for Straumann RC because a material mix-up meant titanium abutments were manufactured from TAN bar material instead of Titanium Grade 4.

    Product
    Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;
    Category
    Distribution
    10 states