The Recall Desk

FDA (Devices) recall action 98464

Merit Medical Systems, Inc. recalled 5 products on 2026-04-01

The FDA (Devices) publishes a recall like this one as 5 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 5 pages.

Products
5
Announced
2026-04-01
Critical
0
Units
472,730

Why these products were recalled

16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for future vascular access.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–5 of 5

  • SevereFDA (Devices)··2026-04-01

    DuraMax hemodialysis catheter sheath introducer may not split as intended

    Merit Medical is recalling DuraMax Chronic Hemodialysis Catheters because a design defect in the 16F dual-valved splittable sheath introducer may prevent it splitting as intended.

    Product
    DuraMax Chronic Hemodialysis Catheter, REF: H787103028015/A, H787103028025/A, H787103028031/A, H787103028035/A, H787103028041/A, H787103028045/A, H787103028051/A, H787103028055/A, H787103028061/A, H787103028075/A, H787103028085/A, H787103028095/A, H787103028165/A…
    Category
    Distribution
    43 states
  • SevereFDA (Devices)··2026-04-01

    Merit 16F splittable sheath introducer recalled as Class I bulk product

    Merit Medical is recalling the 16F Dual Valved Splittable Sheath Introducer in bulk non-sterile form because a design defect may prevent it splitting as intended.

    Product
    16F Dual Valved Splittable Sheath Introducer (bulk, non-sterile), REF: FCL-174-00/B
    Category
    Distribution
    43 states
  • SevereFDA (Devices)··2026-04-01

    CentrosFLO hemodialysis catheters recalled as Class I over sheath introducer defect

    Merit Medical is recalling CentrosFLO Hemodialysis Catheters because a design defect in the 16F dual-valved splittable sheath introducer may prevent it splitting as intended.

    Product
    CentrosFLO Hemodialysis Catheters, REF: CENFP15K/A, CENFP15K/B, CENFP17K/A, CENFP19K/A, CENFP23K/A, CENFP27K/A, CENFP31K/A, CENFT15K, CENFT15K/D, CENFT17K, CENFT17K/A, CENFT17K/D, CENFT19K, CENFT19K/A, CENFT19K/B, CENFT19K/D, CENFT23K, CENFT23K/A, CENFT23K/B, CENFT23K/D…
    Category
    Distribution
    43 states
  • SevereFDA (Devices)··2026-04-01

    ProGuide chronic dialysis catheters recalled as Class I over sheath introducer defect

    Merit Medical is recalling ProGuide Chronic Dialysis Catheters because a design defect in the 16F dual-valved splittable sheath introducer may prevent it splitting as intended.

    Product
    ProGuide Chronic Dialysis Catheters, REF: DC01455550/C, DC21452419/C, DC21452419-NE5/C, DC21452823-NE5/C, DC21454035/C, DC21455550/C
    Category
    Distribution
    43 states
  • SevereFDA (Devices)··2026-04-01

    BioFlo DuraMax catheters recalled as Class I over sheath introducer defect

    Merit Medical is recalling BioFlo DuraMax Catheters because a design defect in the 16F dual-valved splittable sheath introducer may prevent it splitting as intended.

    Product
    BioFlo DuraMax Catheter, REF: H965103028011/A, H965103028021/A, H965103028021/EU, H965103028031/A, H965103028031/EU, H965103028041/A, H965103028041/EU, H965103028051/A, H965103028051/EU, H965103028061/A, H965103028061/EU, H965103028070/A, H965103028070/EU, H965103028080/A…
    Category
    Distribution
    43 states