FDA (Devices) recall action 98346

Beckman Coulter Mishima K.K. recalled 2 products on 2026-03-18

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2026-03-18
Critical
0
Units
189

Why these products were recalled

A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment

A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2026-03-18

    Beckman Coulter AU5800 analyzer may stop analyzing after database errors

    Beckman Coulter is recalling Chemistry Analyzer AU5800 units because leaving the calibration monitor open beyond 15 minutes can trigger database errors that stop the system analyzing.

    Product
    CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281
    Category
    Distribution
    37 states
  • HighFDA (Devices)··2026-03-18

    Beckman Coulter DxC 700 AU analyzer may stop analyzing after errors

    Beckman Coulter is recalling DxC 700 AU analyzers because leaving the calibration monitor open beyond 15 minutes can trigger database errors that stop the system analyzing.

    Product
    DxC 700 AU, REF: B86444, B86446
    Category
    Distribution
    37 states