The Recall Desk

FDA (Devices) recall action 98197

Medline Industries, LP recalled 6 products on 2026-01-21

The FDA (Devices) publishes a recall like this one as 6 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 6 pages.

Products
6
Announced
2026-01-21
Critical
0
Units
60,117

Why these products were recalled

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–6 of 6

  • ModerateFDA (Devices)··2026-01-21

    Laparoscopy kits recalled over suction irrigators that can leak during use

    Medline is recalling 41,432 laparoscopy convenience kits containing Stryker Strykeflow 2 Suction Irrigators, which can leak or emit vaporized saline during use.

    Product
    Medline medical procedure convenience kits labeled as: 1) LAPAROSCOPY CDS-LF, REF CDS860235L; 2) GENERAL LAPAROSCOPIC, REF CDS930223M; 3) LAP CHOLE CDS, REF CDS983678K; 4) LAPAROSCOPY PELV CDS, REF CDS983679K; 5) LAP CHOLE CDS-LF, REF CDS983752M; 6) LAP CHOLE…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2026-01-21

    PALACOS MV bone cement recalled because ampoules may not break

    Heraeus Medical is recalling PALACOS MV pro 40 and PALACOS MV pro 80 bone cement because an unfortunate combination of factors can leave ampoules unbroken, so no appropriate cement dough is formed.

    Product
    Medline medical procedure convenience kits labeled as: 1) WCH BARI-GASTRECTOMY CDS, REF CDS860237AF; 2) ROBOTIC PROCEDURE, REF CDS982128Q; 3) CUH THORACIC, REF CDS984338M; 4) THORACIC PROCEDURE, REF CDS984640I; 5) DAVINCI ROBOTIC PACK-LF, REF DYNJ0525321O; 6) LAP…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2026-01-21

    Robotic and gynaecology kits recalled over leaking suction irrigators inside

    Medline is recalling 12,584 procedure convenience kits containing Stryker Strykeflow 2 Suction Irrigators, which can leak or emit vaporized saline during use.

    Product
    Medline medical procedure convenience kits labeled as: 1) ROBOTIC, REF CDS980523M; 2) ROBOTIC GYN, REF CDS980647K; 3) DAVINCI PROCEDURE, REF CDS980864Q; 4) GYN-URO LAPSCP CDS-LF, REF CDS981583U; 5) GYN-URO LAPSCP CDS-LF, REF CDS981583V; 6) GYN ROBOTIC, REF…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2026-01-21

    Hysterectomy procedure kits recalled over leaking suction irrigators inside

    Medline is recalling 1457 units of procedure convenience kits containing Stryker Strykeflow 2 Suction Irrigators, which can leak or emit vaporized saline during use.

    Product
    Medline medical procedure convenience kits labeled as: 1) LAVH PROCEDURE, REF CDS980754T; 2) GYN LAP HYSTERECTOMY CDS, REF CDS982662Q; 3) DAVINCI HYSTERECTOMY CDS, REF CDS985344F; 4) LAVH, REF DYNJQ9041R; 5) LAPAROSCOPIC HYSTERECTOMY, REF DYNJ900244I; 6) ROBOTIC…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2026-01-21

    Robotic thoracic and cardiac kits recalled over leaking suction irrigators

    Medline is recalling 96 units of procedure convenience kits containing Stryker Strykeflow 2 Suction Irrigators, which can leak or emit vaporized saline during use.

    Product
    Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B; 2) CARDIAC ROBOTIC WHC KIT-LF, REF DYNJ911359
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2026-01-21

    Robotic prostate kits recalled over suction irrigators that can leak

    Medline is recalling 12 units of procedure convenience kits containing Stryker Strykeflow 2 Suction Irrigators, which can leak or emit vaporized saline during use.

    Product
    Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G
    Category
    Distribution
    0 states