ETAC A/S recalled 4 products on 2026-02-11

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2026-02-11
Critical
0
Units
255

Why these products were recalled

A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–4 of 4

  • ModerateFDA (Devices)··2026-02-11

    Hoist sling bars recalled after a lifting bar mounting bolt broke

    ETAC A/S is recalling 64 units of Molift 2-point sling bars, accessories to mobile hoists after a customer reported the bolt for mounting the lifting bar breaking during assembly and load testing.

    Product
    Brand Name: Molift Product Name: Molift 2-point sling bars (Accessory to mobile hoists) Model/Catalog Numbers: (1) 1530101, (2) 1530102 Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due…
    Category
    Distribution
    1 state
  • ModerateFDA (Devices)··2026-02-11

    Molift Mover 205 hoists recalled after a mounting bolt broke

    ETAC A/S is recalling 95 Molift Mover 205 mobile hoists after a customer reported the bolt for mounting the lifting bar breaking during assembly and load testing.

    Product
    Brand Name: Molift Product Name: Molift Mover 205 Model/Catalog Number: (1) M14000, (2) M1401, (3) M1402, (4) M14050, (5) M14150, (6) M1402-01, (7) M14050-00 (8) M14001;
    Category
    Distribution
    1 state
  • ModerateFDA (Devices)··2026-02-11

    Four point sling bars recalled after a mounting bolt broke

    ETAC A/S is recalling 25 Molift 4-point sling bars after a customer reported the bolt for mounting the lifting bar breaking during assembly and load testing.

    Product
    Brand Name: Molift Product Name: Molift 4-point sling bars (Accessory to mobile hoists) Model/Catalog Number: 1430116,1430117,1430115 Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to…
    Category
    Distribution
    1 state
  • ModerateFDA (Devices)··2026-02-11

    Mobile hoists recalled after a lifting bar mounting bolt broke

    ETAC A/S is recalling 71 units of Molift Mover 300 mobile hoists after a customer reported the bolt for mounting the lifting bar breaking during assembly and load testing.

    Product
    Brand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M1501, (2) M1502, (3) M1504, (4) M15050, (5) M15151, (6) M15001, (7) M15002; Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a…
    Category
    Distribution
    1 state