The Recall Desk

FDA (Devices) recall action 98078

Heraeus Medical GmbH (Dental Division) recalled 4 products on 2026-01-21

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2026-01-21
Critical
0
Units
—

Why these products were recalled

increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.

The agency’s own wording, quoted from the enforcement report.

1–4 of 4

  • ModerateFDA (Devices)··2026-01-21

    Procedure kits recalled over suction irrigators that can leak during use

    Medline is recalling 4,536 procedure convenience kits containing Stryker Strykeflow 2 Suction Irrigators, which can leak or emit vaporized saline during use.

    Product
    Brand Name: PALACOS Product Name: PALACOS MV pro 40; PALACOS MV pro 80 Model/Catalog Number: 5150633; 5150635 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2026-01-21

    Swan rubbing alcohol recalled over insanitary conditions at a distribution center

    Gold Star Distribution is recalling Swan White Clear Alcohol over CGMP deviations, citing insanitary conditions including rodent exposure and activity at its distribution center.

    Product
    Brand Name: PALACOS Product Name: PALACOS MV+G pro 40; PALACOS MV+G pro 80 Model/Catalog Number: 5081290; 5081291 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2026-01-21

    PALACOS R bone cement recalled because ampoules may not break

    Heraeus Medical is recalling PALACOS R pro 40 and PALACOS R pro 80 bone cement because a combination of factors can leave ampoules unbroken, so no appropriate cement dough is formed.

    Product
    Brand Name: PALACOS Product Name: PALACOS R pro 40; PALACOS R pro 80 Model/Catalog Number: 5081286; 5081287 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2026-01-21

    PALACOS R plus G bone cement recalled because ampoules may not break

    Heraeus Medical is recalling PALACOS R+G pro 40 and PALACOS R+G pro 80 bone cement because an unfortunate combination of factors can leave ampoules unbroken, so no appropriate cement dough is formed.

    Product
    Brand Name: PALACOS Product Name: PALACOS R+G pro 40; PALACOS R+G pro 80 Model/Catalog Number: 5081288; 5081289 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system…
    Category
    Distribution
    0 states