Why these products were recalled

An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2025-11-12

    Chemistry analyzers recalled over gantry defect that pauses the instrument

    Beckman Coulter is recalling 657 UniCel DxI 800 analyzers because defective gantry sub-components can cause a z-axis motion error that halts the analyzer and delays results.

    Product
    Beckman Coulter UniCel DxI 800, Part Numbers: 973100, A71456; Photometric Chemistry Analyzer
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-11-12

    Immunoassay analyzers recalled over gantry defect that pauses the instrument

    Beckman Coulter is recalling 80 UniCel DxI 600 analyzers because defective gantry sub-components can cause a z-axis motion error that halts the analyzer and delays results.

    Product
    Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer
    Category