The Recall Desk

FDA (Devices) recall action 97496

Medline Industries, LP recalled 5 products on 2025-10-08

The FDA (Devices) publishes a recall like this one as 5 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 5 pages.

Products
5
Announced
2025-10-08
Critical
5
Units
39,565

Why these products were recalled

Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–5 of 5

  • CriticalFDA (Devices)··2025-10-08

    Reprocessed cardiac imaging catheters recalled over residual particulates on device surfaces

    Medline is recalling 7,913 reprocessed ViewFlex Xtra ICE catheters because specific lots may carry small residual particulates on patient-contacting surfaces.

    Product
    Reprocessed ViewFlex Xtra ICE Catheter to visualize cardiac structures; blood flow and other devices within the heart. Medline Item Numbers D087031RH
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2025-10-08

    Reprocessed ultrasound catheters recalled over residual particulates on patient contacting surfaces

    Medline is recalling 7,913 reprocessed ACUSON AcuNav ultrasound catheters because specific lots may carry small residual particulates on patient-contacting surfaces.

    Product
    Reprocessed ACUSON AcuNav Ultrasound Catheter for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. For imaging guidance only, not treatment delivery, during cardiac interventio
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2025-10-08

    Reprocessed electrophysiology catheters recalled over residual particulates on patient surfaces

    Medline is recalling 7,913 reprocessed Inquiry Steerable Diagnostic Catheters because specific lots may carry small residual particulates on patient-contacting surfaces.

    Product
    Reprocessed Inquiry Steerable Diagnostics Catheter for electrogram recording and cardiac stimulation during diagnostic electrophysiologic studies; Medline Item Numbers 81102RH 81104RH 81107RH 81174RH 81402RH 81404RH 81405RH 81472RH 81473RH 81474RH 81531RH 81532RH
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2025-10-08

    Reprocessed steerable electrophysiology catheters recalled over residual particulates on surfaces

    Medline is recalling 7,913 reprocessed Livewire Steerable Electrophysiology Catheters because specific lots may carry small residual particulates on patient-contacting surfaces.

    Product
    Reprocessed Livewire Steerable Electrophysiology Catheter Item Numbers 401575RH 401582RH 401600RH 401603RH 401606RH 401652RH 401653RH 401654RH 401904RH 401905RH 401908RH 401914RH 401915RH 401918RH 401932RH 401933RH 401934RH 401938RH 401939RH 401940RH 401941
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2025-10-08

    Reprocessed cardiac mapping catheters recalled over residual particulates on device surfaces

    Medline is recalling 7,913 reprocessed Webster CS 10-electrode catheters because specific lots may carry small residual particulates on patient-contacting surfaces.

    Product
    Reprocessed Webster CS Catheter 10 Electrodes for electrophysiological mapping of cardiac structures. Medline Item Numbers BD710DF282CRH BD710DF282RRH BD710FJ282CRH BD710FJ282RRH
    Category
    Distribution
    0 states