The Recall Desk

FDA (Devices) recall action 97454

GE Medical Systems, LLC recalled 2 products on 2025-09-03

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2025-09-03
Critical
0
Units
—

Why these products were recalled

GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.

The agency’s own wording, quoted from the enforcement report.

1–2 of 2