The Recall Desk

FDA (Devices) recall action 97396

Boston Scientific Corporation recalled 4 products on 2025-09-17

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2025-09-17
Critical
0
Units
109,913

Why these products were recalled

The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • HighFDA (Devices)··2025-09-17

    UroMax Ultra kits recalled over particles migrating from the inflation device

    Boston Scientific UroMax Ultra Kits are recalled because foreign material particles could migrate from the Encore 26 Inflation Device into a balloon dilatation catheter. 1,369 units are affected.

    Product
    Boston Scientific UroMax Ultra Kit, Material Numbers (UPN): 1. M0062251200, 2. M0062251210, 3. M0062251220, 4. M0062251230, 5. M0062251240, 6. M0062251260, 7. M0062251290, 8. M0062251300, 9. M0062251310. 10. M0062251350, 11. M0062251360, 12. M0062251370.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-09-17

    Encore 26 inflation devices recalled over foreign material particles migrating into catheters

    Boston Scientific Encore 26 Inflation Devices are recalled because foreign material particles could migrate from the device into a balloon dilatation catheter during use. 39,739 units are affected.

    Product
    Boston Scientific Encore 26 Inflation Device, Material Numbers (UPN): 1. H74904526011; 2. H74904526052; 3. M0067101140; 4. M001151050; 5. M001151062; 6. M00566670. (Interventional Cardiology, Peripheral Interventions)
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-09-17

    Encore Advantage kits recalled over particles migrating from the inflation device

    Boston Scientific Encore Advantage Kits are recalled because foreign material particles could migrate from the Encore 26 Inflation Device into a balloon dilatation catheter. 68,347 units are affected.

    Product
    Boston Scientific Encore Advantage Kit, Material Numbers (UPN): 1. H74904527011; 2. H74904527052. (Interventional Cardiology and Peripheral Interventions)
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-09-17

    NephroMax kits recalled because inflation device particles may enter balloon catheters

    Boston Scientific NephroMax Kits are recalled because foreign material particles could migrate from the Encore 26 Inflation Device into a balloon dilatation catheter. 458 units are affected.

    Product
    Boston Scientific NephroMax Kit, Material Numbers (UPN): 1. M0062101180; 2. M0062101600. (Urology)
    Category
    Distribution
    0 states