The Recall Desk

FDA (Devices) recall action 97318

Boston Scientific Corporation recalled 3 products on 2025-08-27

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2025-08-27
Critical
3
Units
—

Why these products were recalled

Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.

The agency’s own wording, quoted from the enforcement report.

1–3 of 3

  • CriticalFDA (Devices)··2025-08-27

    RELIANCE 4-FRONT defibrillation leads recalled over rising LVSI measurements and coil calcification

    RELIANCE 4-FRONT defibrillation leads are recalled because affected models may show gradually rising LVSI measurements linked to calcification of the defibrillation coils, which may lead to early lead replacement.

    Product
    RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658, 0682, 0683, 0685, 0686, 0692, 0693, 0695, 0696
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2025-08-27

    ENDOTAK RELIANCE defibrillation leads recalled over coil calcification and rising LVSI

    ENDOTAK RELIANCE G/SG transvenous defibrillation leads are recalled because affected models may show gradually rising LVSI measurements linked to calcification of the defibrillation coils.

    Product
    ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. 0160, 0161, 0162, 0163, 0164, 0165, 0166, 0167, 0170, 0171, o182, 0173, 0174, 0175, 0176, 0177, 0180, 0182, 0183, 0184, 0185, 0186, 0187
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2025-08-27

    ENDOTAK RELIANCE leads with 4-SITE connector recalled over coil calcification

    ENDOTAK RELIANCE G/SG defibrillation leads with the 4-SITE connector are recalled because affected models may show gradually rising LVSI measurements linked to calcification of the defibrillation coils.

    Product
    ENDOTAK RELIANCE G/SG with 4-SITE Connector, transvenous defibrillation lead, Permanent Defibrillator Electrodes; Model Nos. 0282, 0283, 0285, 0286, 0292, 0293, 0295, 0296
    Category
    Distribution
    0 states