The Recall Desk

FDA (Devices) recall action 97287

Diagnostica Stago, Inc. recalled 4 products on 2025-09-03

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2025-09-03
Critical
0
Units

Why these products were recalled

Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–4 of 4

  • ModerateFDA (Devices)··2025-09-03

    Fibrinogen test kits recalled over potential cross-contamination in heparin assays

    STA-Fibrinogen 5 kits are recalled because of the potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in heparin applications. 829,800 units are affected, 245,200 of them in the US.

    Product
    STA - Fibrinogen 5. Model/Catalog Number: 00674. The STA - Fibrinogen 5 kit is intended for use with the analyzers of the STA-R, STA Compact and STA Satellite families for the quantitative determination of fibrinogen levels in human citrated plasma by the clotting method of
    Category
    Distribution
    22 states
  • ModerateFDA (Devices)··2025-09-03

    Anti-Xa 4 test kits recalled over potential fibrinogen reagent contamination

    STA-Liquid Anti-Xa 4 kits are recalled because of the potential for contamination by Stago Fibrinogen reagents in unfractionated and low molecular weight heparin applications. 116,600 units are affected.

    Product
    STA-Liquid Anti-Xa 4. Model/Catalog Number: 00311US. The STA - Liquid Anti-Xa kits are intended for use with STA-R, STA Compact and STA Satellite analyzers, for the quantitative determination of the plasma levels of unfractionated (UFH) and low molecular weight (LMWH) heparins
    Category
    Distribution
    22 states
  • ModerateFDA (Devices)··2025-09-03

    Anti-Xa test kits recalled over potential contamination by fibrinogen reagents

    STA-Liquid Anti-Xa 8 kits are recalled because of the potential for contamination by Stago Fibrinogen reagents in unfractionated and low molecular weight heparin applications. 126,200 units are affected.

    Product
    STA-Liquid Anti-Xa 8. Model/Catalog Number: 00322US. The STA - Liquid Anti-Xa kits are intended for use with STA-R, STA Compact and STA Satellite analyzers, for the quantitative determination of the plasma levels of unfractionated (UFH) and low molecular weight (LMWH) heparin
    Category
    Distribution
    22 states
  • ModerateFDA (Devices)··2025-09-03

    Coagulation analyzers recalled over potential reagent contamination in anti-Xa testing

    STA Satellite analyzers are recalled because of the potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in unfractionated and low molecular weight heparin applications. 3,650 units are affected.

    Product
    STA Satellite. Product Code: All references. All software versions. The STA Satellite is a fully automated benchtop analyzer capable of simultaneously performing clotting, chromogenic, and immunologic assays.
    Category
    Distribution
    22 states