The Recall Desk

FDA (Devices) recall action 97273

C.R. Bard Inc recalled 3 products on 2025-09-24

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2025-09-24
Critical
0
Units
12,270

Why these products were recalled

Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with infection control coating, when it's silicone hydrogel with antimicrobial coating, or 3)Natural latex with infection control, when it's silicone hydrogel coated without antimicrobial coating, which may cause infection and/or local and/or systemic allergic reaction

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • HighFDA (Devices)··2025-09-24

    Bardex Foley catheter trays recalled over inserts describing the wrong material

    SureStep Foley Tray Systems, REF A303316A lot NGJX6399, are recalled because incorrect inserts may lead users to misidentify the catheter material or coating, risking infection or allergic reaction.

    Product
    SureStep Foley Tray System Bardex I.C. Complete Care Infection Control Foley Catheter Tray, REF: A303316A
    Category
    Distribution
    33 states
  • HighFDA (Devices)··2025-09-24

    Lubri-Sil Foley trays recalled over inserts describing the wrong catheter material

    SureStep Foley Tray Systems with Bard Lubri-Sil catheters, REF A942216 lot NGJX4320, are recalled because incorrect inserts may lead users to misidentify the catheter material or coating. 3,670 units are affected.

    Product
    SureStep Foley Tray System BARD¿ Lubri-Sil Foley Catheter Tray, REF: A942216
    Category
    Distribution
    33 states
  • HighFDA (Devices)··2025-09-24

    Foley catheter trays recalled because inserts describe the wrong catheter material

    SureStep Foley Tray Systems, REF A303414A lot NGJY3685, are recalled because incorrect inserts may lead users to misidentify the catheter material or coating, risking infection or allergic reaction.

    Product
    SureStep Foley Tray System Lubri-Sil I.C. Complete Care Infection Control Foley Catheter Tray, REF: A303414A
    Category
    Distribution
    33 states