PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalled 2 products on 2025-09-10

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2025-09-10
Critical
0
Units
2,867

Why these products were recalled

Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2025-09-10

    Azurion imaging systems recalled because single phase UPS may fail on power loss

    Philips Azurion systems configured with a 1 Phase UPS are recalled because they may shut down and/or fail to start up if power to the system is lost. 1,861 units are affected.

    Product
    Philips Azurion System configured with a 1 Phase UPS. Azurion family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Azurion 3 M12, Model Number 722063. 2. Azurion 3 M15, Model Number 722064. 3. Azurion 7 B12, Model Number 722067. 4.…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-09-10

    Interventional imaging systems recalled because single phase UPS may fail on power loss

    Philips Allura systems configured with a 1 Phase UPS are recalled because they may shut down and/or fail to start up if power to the system is lost. 1,006 units are affected.

    Product
    Philips Allura System configured with a 1 Phase UPS. Allura family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003 2. Allura Xper FD10/10, Model Number 722005 3. Allura Xper FD20, Model Number 722006…