The Recall Desk

FDA (Devices) recall action 97172

Philips North America recalled 19 products on 2025-08-06

The FDA (Devices) publishes a recall like this one as 19 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 19 pages.

Products
19
Announced
2025-08-06
Critical
0
Units
8,052

Why these products were recalled

The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–19 of 19

  • HighFDA (Devices)··2025-08-06

    Philips Achieva 3.0T gradient coils may act as a heat source

    Achieva 3.0T MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 460 units are affected.

    Product
    Achieva 3.0T; Model Numbers (REF): (1) 781177, (2) 781277, (3) 781278, (4) 781344, (5) 781345;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-08-06

    Philips MRI Machine Intera 3.0T Quasar Dual Recalled for Fire Risk

    Philips is recalling its Intera 3.0T Quasar Dual MRI system due to a potential fire hazard from gradient coil failures. The recall affects one unit distributed worldwide.

    Product
    Intera 3.0T Quasar Dual Model Number (REF): 781150;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-08-06

    Philips Ingenia 1.5T gradient coils may act as a heat source

    Ingenia 1.5T MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 2,279 units are affected.

    Product
    Ingenia 1.5T Model Numbers (REF): (1) 781315, (2) 718641, (3) 781396, (4) 782101, (5) 782115, (6) 782140;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-08-06

    Philips Achieva XR gradient coils may act as a heat source

    Achieva XR MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 40 units are affected.

    Product
    Achieva XR; Model Numbers (REF): (1) 781153, (2) 781253;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-08-06

    SmartPath to Ingenia Elition X gradient coils may overheat

    SmartPath to Ingenia Elition X systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 33 units affected.

    Product
    SmartPath to Ingenia Elition X Model Numbers (REF): (1) 782118, (2) 782144;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-08-06

    Philips Ingenia 3.0T gradient coils may act as a heat source

    Ingenia 3.0T MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 1,467 units are affected.

    Product
    Ingenia 3.0T Model Numbers (REF): (1) 781342, (2) 781377, (3) 782103;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-08-06

    Philips Ingenia Ambition S gradient coils may act as heat source

    Ingenia Ambition S MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 611 units are affected.

    Product
    Ingenia Ambition S Model Numbers (REF): (1) 781359, (2) 782108, (3) 782133 (China ONLY), (4) 782139;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-08-06

    Philips Ingenia 3.0T CX gradient coils may act as heat source

    Ingenia 3.0T CX MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 350 units are affected.

    Product
    Ingenia 3.0T CX Model Numbers (REF): (1) 781271, (2) 782105;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-08-06

    Philips MR 7700 gradient coils may act as a heat source

    MR 7700 systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 82 units are affected.

    Product
    MR 7700 Model Numbers (REF): (1) 782120, (2) 782153;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-08-06

    SmartPath to dStream XR gradient coils may act as heat source

    SmartPath to dStream for XR and 3.0T systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 294 units affected.

    Product
    SmartPath to dStream for XR and 3.0T Model Numbers (REF): (1) 781270, (2) 782113, (3) 782129;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-08-06

    Philips Ingenia Ambition X gradient coils may act as heat source

    Ingenia Ambition X MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 690 units are affected.

    Product
    Ingenia Ambition X Model Numbers (REF): (1) 781356, (2) 782109, (3) 782138;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-08-06

    Philips Evolution Upgrade 3.0T gradient coils may act as heat source

    Evolution Upgrade 3.0T MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 50 units are affected.

    Product
    Evolution Upgrade 3.0T Model Numbers (REF): (1) 782117, (2) 782143;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-08-06

    Philips Ingenia Elition S gradient coils may act as a heat source

    Ingenia Elition S MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 242 units are affected.

    Product
    Ingenia Elition S Model Numbers (REF): (1) 781357, (2) 782106, (3) 782137, (4) 782150 (China ONLY);
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-08-06

    Philips Evolution Upgrade 1.5T gradient coils may act as a heat source

    Evolution Upgrade 1.5T MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 93 units are affected.

    Product
    Evolution Upgrade 1.5T Model Numbers (REF): (1) 782116, (2) 782148;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-08-06

    Philips Ingenia Elition X gradient coils may act as heat source

    Ingenia Elition X MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 632 units are affected.

    Product
    Ingenia Elition X Model Numbers (REF): (1) 781358, (2) 782107, (3) 782119, (4) 782136, (5) 782151 (China ONLY);
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-08-06

    SmartPath to dStream 3.0T gradient coils may act as heat source

    SmartPath to dStream for 3.0T systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 15 units affected.

    Product
    SmartPath to dStream for 3.0T Model Number (REF): 782145;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-08-06

    Philips Ingenia 1.5T S gradient coils may act as heat source

    Ingenia 1.5T S MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 136 units are affected.

    Product
    Ingenia 1.5T S Model Number (REF): 781347;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-08-06

    Philips MR 7700 upgrade gradient coils may act as a heat source

    Upgrade to MR 7700 systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 26 units are affected.

    Product
    Upgrade to MR 7700 Model Number (REF): 782130;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-08-06

    Philips MR 5300 gradient coils may act as a heat source

    MR 5300 systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 551 units are affected.

    Product
    MR 5300 Model Numbers (REF): (1) 782110, (2) 782135 (China ONLY), (3) 782152;
    Category
    Distribution
    0 states