The Recall Desk

FDA (Devices) recall action 97047

Ortho-Clinical Diagnostics, Inc. recalled 4 products on 2025-07-23

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2025-07-23
Critical
0
Units
414,000

Why these products were recalled

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous pat…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–4 of 4

  • HighFDA (Devices)··2025-07-23

    VITROS Performance Verifier II lots can cause quality control drift

    VITROS Chemistry Products Performance Verifier II is being recalled because quality control drift can occur with certain cholinesterase slide lots, which could produce erroneous patient results.

    Product
    Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier II Model/Catalog Number: 8231474 Software Version: N/A Product Description: VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on…
    Category
    Distribution
    13 states
  • HighFDA (Devices)··2025-07-23

    VITROS Performance Verifier I lots can cause quality control drift

    VITROS Chemistry Products Performance Verifier I is being recalled because quality control drift can occur with certain cholinesterase slide lots, which could produce erroneous patient results.

    Product
    Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier I Model/Catalog Number: 806 7324 Software Version: Not Applicable Product Description: VITROS Chemistry Products Performance Verifier is an assayed control used to monitor…
    Category
    Distribution
    13 states
  • HighFDA (Devices)··2025-07-23

    VITROS CHE slides sold in Japan can cause quality control drift

    VITROS Chemistry Products CHE Slides supplied in Japan are being recalled because quality control drift can occur with certain Performance Verifier lots, which could produce erroneous patient results.

    Product
    VITROS Chemistry Products CHE Slides (Japan); Catalog Numbers: (1) 684 4074 (90 slides), (2) 684 4138 (300 slides);
    Category
    Distribution
    13 states
  • HighFDA (Devices)··2025-07-23

    VITROS CHE slides can cause quality control drift with performance verifiers

    VITROS Chemistry Products CHE Slides are being recalled because quality control drift can occur with certain Performance Verifier lots, which could produce erroneous patient results.

    Product
    VITROS Chemistry Products CHE Slides; Catalog Numbers: (1) 191 4605 (300 slides), (2) 800 4707 (90 slides);
    Category
    Distribution
    13 states