The Recall Desk

FDA (Devices) recall action 96999

Philips North America recalled 9 products on 2025-06-18

The FDA (Devices) publishes a recall like this one as 9 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 9 pages.

Products
9
Announced
2025-06-18
Critical
0
Units
7

Why these products were recalled

identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or even death. This issue could also lead to property damage.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–9 of 9

  • SevereFDA (Devices)··2025-06-18

    Philips Intera Achieva 1.5T Pulsar MRI Systems Gradient Coil Fire Risk

    Philips is recalling certain Intera Achieva 1.5T Pulsar MRI systems due to a potential component failure in the gradient coil that could generate smoke and/or fire, risking patient and operator injury or death.

    Product
    Intera Achieva 1.5T Pulsar; Product Number: 781171;
    Category
    Distribution
    8 states
  • SevereFDA (Devices)··2025-06-18

    Philips Intera 1.5T Achieva Nova MRI Systems Gradient Coil Fire Risk

    Philips is recalling certain Intera 1.5T Achieva Nova MRI systems due to a gradient coil component that can overheat, potentially causing smoke or fire with risk of burns, smoke inhalation, or asphyxia.

    Product
    Intera 1.5T Achieva Nova Product Number: 781172;
    Category
    Distribution
    8 states
  • SevereFDA (Devices)··2025-06-18

    Philips Intera 1.5T Achieva Nova-Dual MRI System Fire Hazard Recall

    Philips is recalling three Intera 1.5T Achieva Nova-Dual MRI systems worldwide due to a gradient coil component failure that can generate smoke and fire, posing risks including burn injury and asphyxia.

    Product
    Intera 1.5T Achieva Nova-Dual Product Number: 781173;
    Category
    Distribution
    8 states
  • HighFDA (Devices)··2025-06-18

    Philips Intera 1.5T MRI Systems Gradient Coil Fire Risk Recall

    Philips is recalling two Intera 1.5T MRI systems due to a gradient coil component failure that can produce smoke or fire, posing risks of inhalation injury, burns, asphyxia, or death.

    Product
    Intera 1.5T Master/Nova Product Number: 781106;
    Category
    Distribution
    8 states
  • SevereFDA (Devices)··2025-06-18

    Philips Intera 1.5T MRI Systems Gradient Coil Fire Risk Recall

    Philips is recalling Intera 1.5T Omni/Stellar MRI systems due to a gradient coil component failure that can produce smoke and fire, risking patient and operator injury or death.

    Product
    Intera 1.5T Omni/Stellar; Product Number: 781104;
    Category
    Distribution
    8 states
  • SevereFDA (Devices)··2025-06-18

    Philips Achieva XR MRI systems recall for fire hazard

    Philips Achieva XR MRI systems may have a gradient coil component failure that can generate heat, smoke, and fire. The risk to patients and operators includes inhalation of smoke, burns, and asphyxia.

    Product
    Achieva XR Product Numbers: (1) 781153, (2) 781253;
    Category
    Distribution
    8 states
  • SevereFDA (Devices)··2025-06-18

    Philips Intera 1.5T Explorer/Nova Dual MRI Gradient Coil Fire Risk

    Philips is recalling one Intera 1.5T Explorer/Nova Dual MRI system due to a gradient coil component that may overheat, produce smoke, or ignite, risking patient and operator injury.

    Product
    Intera 1.5T Explorer/Nova Dual Product Number: 781108;
    Category
    Distribution
    8 states
  • SevereFDA (Devices)··2025-06-18

    Philips SmartPath to dStream 1.5T MRI Gradient Coil Fire Risk

    Philips SmartPath to dStream for 1.5T MRI systems are recalled due to potential gradient coil component failure that may produce smoke and fire, posing risks of inhalation injury, burns, asphyxia, and death.

    Product
    SmartPath to dStream for 1.5T¿; Product Number: 782146;
    Category
    Distribution
    8 states
  • SevereFDA (Devices)··2025-06-18

    Philips Intera 1.5T Achieva IT Nova MRI System Gradient Coil Fire Risk

    Philips is recalling one Intera 1.5T Achieva IT Nova MRI system due to a gradient coil component that may overheat and produce smoke or fire, posing risks of inhalation injury, burns, or death.

    Product
    Intera 1.5T Achieva IT Nova Product Number: 781175
    Category
    Distribution
    8 states