The Recall Desk

FDA (Devices) recall action 96992

GE Medical Systems China Co., Ltd. recalled 2 products on 2025-07-02

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2025-07-02
Critical
0
Units
826

Why these products were recalled

GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2025-07-02

    GE Healthcare LOGIQ F R3 Ultrasound System Incorrect Fetal Weight Data Display

    GE Healthcare is recalling the LOGIQ F R3 ultrasound system because it may display previous obstetric patient data in fetal weight measurements, potentially leading to incorrect clinical decisions about fetal size and growth.

    Product
    GE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic ultrasound system
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-07-02

    GE Healthcare Versana Ultrasound Systems display incorrect fetal weight data

    GE Healthcare Versana Premier R3 and LOGIQ F R3 ultrasound systems can display previous obstetric patient data in fetal weight measurement reports, potentially leading to incorrect clinical decisions based on inaccurate fetal size and growth calculations.

    Product
    GE Healthcare Versana Ultrasound Systems, sold under the following names and Model/Catalog Numbers: 1) Versana Premier R3 VS, Model/Catalog Number 5938558; 2) Versana Premier R3 VA, Model/Catalog Number 5938559; 3) Versana Premier Lotus 4PP, Model/Catalog Number 5938560…
    Category
    Distribution
    0 states