The Recall Desk

FDA (Devices) recall action 96946

Ortho-Clinical Diagnostics, Inc. recalled 3 products on 2025-07-09

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2025-07-09
Critical
0
Units

Why these products were recalled

a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date is omitted during manual loading. The issue was identified by a customer complaint, in which a CREA Urine test was processed with an RE (Reagent Expired) code, even though the diluent pack was not expired and the use-expired-reagents setting was disabled. This can result in erroneous test results and/or delaying test results.

A software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date is omitted during manual loading. The issue was identified by a customer complaint, in which a CREA Urine test was processed with an RE (Reagent Expired) code, even though the diluent pack was not expired and the use-expired-reagents setting was disabled. This can result in erroneous test results and/or delaying test results.

The agency’s own wording, quoted from the enforcement report.

1–3 of 3

  • HighFDA (Devices)··2025-07-09

    VITROS 4600 chemistry systems may report results using expired diluent packs

    VITROS 4600 Chemistry Systems are being recalled because a software anomaly can allow results to be reported using diluent packs the system treats as expired, which can produce erroneous results.

    Product
    Brand Name: VITROS Product Name: VITROS 4600 Chemistry System Model/Catalog Number: (1) 6802445, (2) 6900440 Software Version: version 3.2 to version 3.8.3 (all available software versions) Product Description: In vitro diagnostic chemistry analyzer.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2025-07-09

    VITROS 5600 Integrated System software anomaly reported

    Ortho-Clinical Diagnostics is recalling the VITROS 5600 Integrated System due to a software anomaly that can report test results using expired reagents even when the use-expired-reagents setting is disabled, potentially causing erroneous or delayed test results.

    Product
    Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: (1) 6802413, (2) 6802915
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-07-09

    VITROS XT 7600 Chemistry Analyzer Software Anomaly Causing Erroneous Test Results

    Ortho-Clinical Diagnostics is recalling VITROS XT 7600 Integrated System analyzers due to a software anomaly that can report test results as expired when reagent shelf expiration dates are omitted during manual loading, potentially causing erroneous or delayed test results.

    Product
    Brand Name: VITROS; Product Name: VITROS XT 7600 Integrated System; Model/Catalog Number: (1) 6844461, (2) 6272222; Software Version: version 3.2 to version 3.8.3 (all available software versions); Product Description: In vitro diagnostic chemistry analyzer.
    Category
    Distribution
    0 states