Beckman Coulter Mishima K.K. recalled 2 products on 2025-06-11

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2025-06-11
Critical
0
Units
16,342

Why these products were recalled

Sample probes with defective probe tips, when used with clinical chemistry analyzers, may lead to a lower concentration being dispensed, which may not be detectable by QC checks, which could lead to the reporting of false low test results (Max 34% error at 1.0 microliters and 7% error at 1.6 microliters sample dispensing volume), which may lead to delayed recognition and treatment.

A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing. The air layer is the air space that separates the cleaning water filled in the probe from the sample sucked into the probe. Maintaining this air space prevents the cleaning water and the sample from mixing. In the defective S probe, the "air layer" was not properly formed, causing the cleaning water and the sample to mix diluting the sample. As a result, the sample mixture dispe…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2025-06-11

    Beckman Coulter Sample Probes with Defective Tips Clinical Chemistry

    Beckman Coulter is recalling sample probes with defective tips used in clinical chemistry analyzers. The defective probes may dispense lower sample concentrations, potentially leading to false low test results and delayed diagnosis.

    Product
    Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 - 179433670
    Category
    Distribution
    53 states
  • ModerateFDA (Devices)··2025-06-11

    Beckman Coulter Sample Probe Manufacturing Defect Recall

    Beckman Coulter Mishima is recalling sample probes due to a manufacturing defect where rounded and chamfered corners prevent proper air layer formation, causing sample dilution and inaccurate test results.

    Product
    Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 - 179433670
    Category
    Distribution
    53 states