FDA (Devices) recall action 96634

GE Medical Systems China Co., Ltd. recalled 15 products on 2025-05-07

The FDA (Devices) publishes a recall like this one as 15 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 15 pages.

Products
15
Announced
2025-05-07
Critical
0
Units
14,448

Why these products were recalled

Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–15 of 15

  • SevereFDA (Devices)··2025-05-07

    GE Healthcare CARESTATION 650C A1 Anesthesia System Ventilation Recall

    GE Healthcare CARESTATION 650C A1 anesthesia systems may fail to provide effective ventilation in Volume Control Ventilation mode. Ventilation can be achieved using alternate modes or manual ventilation.

    Product
    GE Healthcare CARESTATION 650C A1, Model/REF Number 1012-9655-000. Anesthesia system.
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2025-05-07

    GE Healthcare CARESTATION 620 A2 Anesthesia System Ventilation Failure

    Certain GE Healthcare CARESTATION anesthesia systems may fail to provide effective ventilation in Volume Control Ventilation (VCV) mode. Alternative ventilation modes or manual ventilation can be used.

    Product
    GE Healthcare CARESTATION 620 A2, Model/REF Number 1012-9620-002. Anesthesia system.
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2025-05-07

    GE Healthcare CARESTATION anesthesia systems may not ventilate effectively

    Certain GE Healthcare CARESTATION 620, 650, 650c, 750, and 750c anesthesia delivery systems may not provide effective ventilation in Volume Control Ventilation (VCV) mode. Effective ventilation is available using other ventilation modes or manual ventilation.

    Product
    GE Healthcare CARESTATION 650 A2, Model/REF Number 1012-9650-002. Anesthesia system.
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2025-05-07

    GE Healthcare Anesthesia Systems Recalled Over Ventilation Failure

    GE Healthcare is recalling certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems because they may not provide effective ventilation in Volume Control Ventilation (VCV) mode. The issue does not affect Pressure Control Ventilation (PCV), PCV-VG, or manual ventilation modes.

    Product
    GE Healthcare Carestation 650 SE, Model/REF Number 1012-9650-012. Anesthesia system.
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2025-05-07

    GE Healthcare Carestation 620 SE Anesthesia System Ventilation Defect Recall

    GE Healthcare is recalling certain Carestation 620 SE anesthesia systems (27 units) because they may not provide effective ventilation in Volume Control Ventilation mode. Alternative ventilation modes remain functional.

    Product
    GE Healthcare Carestation 620 SE, Model/REF Number 1012-9620-012. Anesthesia system.
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2025-05-07

    GE Healthcare CARESTATION Anesthesia Systems Ventilation Mode Failure Recall

    Certain GE Healthcare CARESTATION 750 A2 anesthesia systems may not provide effective ventilation in Volume Control Ventilation mode. Users should switch to Pressure Control Ventilation or Manual ventilation modes.

    Product
    GE Healthcare CARESTATION 750 A2, Model/REF Number 1012-9750-002. Anesthesia system.
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2025-05-07

    GE Healthcare CARESTATION 750 Anesthesia System Volume Control Ventilation Failure

    Certain GE Healthcare CARESTATION 750 anesthesia systems may fail to provide effective ventilation in Volume Control Ventilation mode. Affected systems can deliver ventilation through alternative modes or manual ventilation.

    Product
    GE Healthcare CARESTATION 750 A1, Model/REF Number 1012-9750-000. Anesthesia system.
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2025-05-07

    GE Healthcare Anesthesia Systems Recalled Over Ventilation Failure

    GE Healthcare is recalling certain Carestation 620/650/650c and 750/750c anesthesia delivery systems because they may not provide effective ventilation in Volume Control Ventilation (VCV) mode. The issue does not affect Pressure Control Ventilation (PCV), PCV-VG, or manual ventilation modes.

    Product
    GE Healthcare Carestation 620 A1, Model/REF Number 1012-9620-200. Anesthesia system.
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2025-05-07

    GE Healthcare Carestation 650 A1 Anesthesia System Ventilation

    Certain GE Healthcare Carestation 650 A1 anesthesia systems may fail to provide effective ventilation in Volume Control Ventilation mode. Alternative ventilation modes or manual ventilation can be used.

    Product
    GE Healthcare CARESTATION 650 A1, Model/REF Number 1012-9650-000. Anesthesia system.
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2025-05-07

    GE Healthcare Carestation 650c A1 Anesthesia System Ventilation Failure

    GE Healthcare is recalling certain Carestation 650c A1 anesthesia systems worldwide because they may not provide effective ventilation in Volume Control Ventilation mode, limiting patient ventilation options during anesthesia delivery.

    Product
    GE Healthcare Carestation 650c A1, Model/REF Number 1012-9655-200. Anesthesia system.
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2025-05-07

    GE Healthcare CARESTATION Anesthesia Systems Volume Control Ventilation Failure

    GE Healthcare CARESTATION anesthesia delivery systems may fail to provide effective ventilation in Volume Control Ventilation mode. Affected units can deliver ventilation through Pressure Control Ventilation modes or manual ventilation.

    Product
    GE Healthcare CARESTATION 650C A2, Model/REF Number 1012-9655-002. Anesthesia system.
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2025-05-07

    GE Healthcare CARESTATION 750c A1 Anesthesia System Ventilation Defect

    GE Healthcare CARESTATION 750c A1 anesthesia systems may not provide effective ventilation in Volume Control Ventilation mode. Affected units can achieve ventilation through alternative modes or manual operation.

    Product
    GE Healthcare CARESTATION 750c A1, Model/REF Number 1012-9755-000. Anesthesia system.
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2025-05-07

    GE Healthcare CARESTATION anesthesia systems ventilation malfunction recall

    GE Healthcare CARESTATION 620 A1 anesthesia systems may fail to provide effective ventilation in Volume Control Ventilation mode. Effective ventilation is available in Pressure Control or Manual ventilation modes.

    Product
    GE Healthcare CARESTATION 620 A1, Model/REF Number 1012-9620-000. anesthesia systems
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2025-05-07

    GE Healthcare Carestation Anesthesia Systems Ventilation Defect Recall

    Certain GE Healthcare Carestation anesthesia delivery systems may fail to provide effective ventilation in Volume Control Ventilation (VCV) mode. Affected systems can still be used safely in Pressure Control Ventilation or Manual ventilation modes.

    Product
    GE Healthcare Carestation 650 A1, Model/REF Number 1012-9650-200. Anesthesia system.
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2025-05-07

    GE Healthcare CARESTATION Anesthesia Systems may not ventilate properly

    GE Healthcare is recalling certain CARESTATION anesthesia systems that may fail to provide effective ventilation in Volume Control Ventilation mode. Alternative ventilation modes remain functional.

    Product
    GE Healthcare CARESTATION 750c A2, Model/REF Number 1012-9755-002. Anesthesia system.
    Category
    Distribution
    1 state