FDA (Devices) recall action 96556

Boston Scientific Corporation recalled 4 products on 2025-05-07

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2025-05-07
Critical
0
Units
777

Why these products were recalled

Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • ModerateFDA (Devices)··2025-05-07

    IceSeed 1.5 CX S Cryoablation Needle Programmed with Incorrect Settings

    Boston Scientific is recalling 435 IceSeed 1.5 CX S cryoablation needles manufactured between April 2024 and February 2025 that were programmed with DEMO settings instead of Commercial settings, which can cause the Cryoablation System to prompt repeated tests if the needle is disconnected and reconnected.

    Product
    IceSeed 1.5 CX S NEEDLE US, Cryoablation Needle, REF H7493967233100. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2025-05-07

    IceSeed 1.5 CX Cryoablation Needles Recalled for Incorrect Software Settings

    Boston Scientific is recalling IceSeed 1.5 CX cryoablation needles manufactured between April 2024 and February 2025 due to incorrect software programming. Affected needles may prompt unnecessary Needle Integrity and Functionality Tests if disconnected and reconnected during use.

    Product
    IceSeed 1.5 CX NEEDLE OUS, Cryoablation Needle, REF H7493967534170. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2025-05-07

    IceSeed CX Cryoablation Needles Recalled for Demo Programming Issue

    Boston Scientific is recalling IceSeed 1.5 CX cryoablation needles manufactured between April 2024 and February 2025 because they were programmed with demo settings instead of commercial settings. If disconnected and reconnected during use, the system will prompt unnecessary recalibration testing.

    Product
    IceSeed 1.5 CX NEEDLE US, Cryoablation Needle, REF H7493967433170. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2025-05-07

    Boston Scientific IceSeed CX Cryoablation Needles Incorrectly Programmed

    Boston Scientific is recalling IceSeed 1.5 CX S cryoablation needles manufactured April 2024 to February 2025 that were programmed with DEMO instead of Commercial settings. If disconnected and reconnected after initial testing, the system will re-prompt testing rather than functioning normally.

    Product
    IceSeed 1.5 CX S NEEDLE OUS, Cryoablation Needle, REF H7493967334100. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.
    Category
    Distribution
    0 states