FDA (Devices) recall action 96539

Siemens Healthcare Diagnostics Inc recalled 2 products on 2025-05-07

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2025-05-07
Critical
0
Units
269,459

Why these products were recalled

The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2025-05-07

    epoc BGEM BUN Test Cards for Blood Analysis System Recalled

    Siemens Healthcare Diagnostics is recalling epoc BGEM BUN Test Cards (25 pk) due to discrepant high pH results when samples are introduced with higher injection volumes on affected sensor configurations.

    Product
    epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515; epoc Software Version & Sensor Configuration: epoc Host SW v3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2025-05-07

    epoc BGEM Crea Test Card for Blood Analysis System recalled

    Siemens Healthcare Diagnostics is recalling epoc BGEM Crea Test Card (25 pk) units due to discrepant high pH results in samples with higher injection volumes. The recall affects 5,207 units distributed nationwide and internationally.

    Product
    epoc BGEM Crea Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736382; epoc Software Version & Sensor Configuration: epoc Host SW v3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n.
    Category
    Distribution
    0 states