Tandem Diabetes Care, Inc. recalled 2 products on 2025-04-23

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2025-04-23
Critical
0
Units
148

Why these products were recalled

A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated Glucose Value (EGV) is above 255 mg/dL at the start or end of a gap in data collection due to a lapse in connection from a paired continuous glucose monitor (CGM) sensor, which can lead to under-delivery or over-delivery of insulin based on inaccurate result leading to severe cases of hypoglycemia or hyperglycemia.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • SevereFDA (Devices)··2025-04-23

    Tandem Mobi Insulin Pump Software Defect Risks Incorrect Insulin Delivery

    Tandem Diabetes Care is recalling Tandem Mobi Insulin Pumps with software version 7.9 due to a defect that can cause incorrect insulin delivery when glucose data gaps occur, potentially leading to severe high or low blood sugar.

    Product
    Tandem Mobi Insulin Pump with Interoperable Technology
    Category
    Distribution
    29 states
  • SevereFDA (Devices)··2025-04-23

    t:slim X2 Insulin Pump Software Defect Affecting Glucose Delivery

    Tandem Diabetes Care is recalling 63 units of the t:slim X2 Insulin Pump with software version 7.9 due to a defect that can cause incorrect insulin delivery. The defect may lead to dangerous blood sugar levels when glucose monitoring loses connection.

    Product
    t:slim X2 Insulin Pump with Interoperable Technology
    Category
    Distribution
    29 states