FDA (Devices) recall action 96387

Checkpoint Surgical Inc recalled 2 products on 2025-04-02

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2025-04-02
Critical
0
Units
587

Why these products were recalled

Adhesive on the back of the lead may not fully encapsulate the metallic portion of the wire and button subassembly, potentially leading to exposed metal. There is a risk of electrical current leakage, which may lead to inconsistent or confusing muscle responses, potentially compromising the effectiveness of the intended use as the surgeon could take an incorrect clinical action.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2025-04-02

    Checkpoint Guardian Intraoperative Lead adhesive defect electrical leakage risk

    The adhesive on Checkpoint Guardian Intraoperative Leads may not fully cover metallic wire components, leading to potential electrical current leakage. This could result in inconsistent muscle responses and affect surgical effectiveness.

    Product
    Brand Name: Checkpoint Guardian Intraoperative Lead Product Name: Checkpoint Guardian Intraoperative Lead, Small Model/Catalog Number: REF# 9524 (Small) Product Description: The Checkpoint Guardian Intraoperative Leads are single-patient disposable accessories for providing…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-04-02

    Checkpoint Guardian Intraoperative Lead: Potential Electrical Exposure Risk

    Adhesive may not fully encapsulate metallic components on Checkpoint Guardian Intraoperative Leads, creating potential for electrical current leakage. This could cause inconsistent muscle responses and incorrect surgical actions.

    Product
    Brand Name: Checkpoint Guardian Intraoperative Lead Product Name: Checkpoint Guardian Intraoperative Lead, Medium Model/Catalog Number: REF# 9525 (Medium) Product Description: The Checkpoint Guardian Intraoperative Leads are single-patient disposable accessories for…
    Category
    Distribution
    0 states