GE Healthcare (China) Co., Ltd. recalled 5 products on 2025-03-05

The FDA (Devices) publishes a recall like this one as 5 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 5 pages.

Products
5
Announced
2025-03-05
Critical
0
Units
3,022

Why these products were recalled

For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–5 of 5

  • HighFDA (Devices)··2025-03-05

    GE Healthcare MRI system recalled for RF energy safety control failure

    GE Healthcare's Optima MR360 MRI systems may allow unsafe RF energy exposure if a scan resumes after power loss. The user interface can incorrectly enable scanning when tissue-heating levels exceed safe limits.

    Product
    Optima MR360, NMRI system
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-03-05

    GE SIGNA MR380 NMRI System SAR Limit Exceeded During Scan Resumption

    GE SIGNA MR380 NMRI systems may allow scans to resume when the 6-minute average SAR (tissue heating rate) exceeds safe limits, potentially causing excessive tissue heating especially in patients with implants or conductive contact.

    Product
    SIGNA MR380, NMRI system, Non-US Only
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-03-05

    MRI Systems May Resume Scans When SAR Exceeds Safe Limits

    Certain GE Healthcare MRI systems may incorrectly enable scan resumption when radiofrequency energy absorption exceeds FDA-set limits, potentially causing tissue heating. Risk is higher for patients with implants or touching conductive materials.

    Product
    SIGNA MR355, SIGNA MR360, NMRI system
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-03-05

    MRI System May Resume Scanning Above Safe Thermal Limits

    GE Healthcare's Brivo MR355 MRI system can incorrectly resume scans when tissue heating limits are exceeded, potentially causing thermal burns if patients contact conductive materials. The recall affects 22 systems distributed nationwide and internationally.

    Product
    Brivo MR355, NMRI system
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-03-05

    MRI Systems Allow Unsafe Scan Resumption Exceeding SAR Safety Limits

    GE Healthcare MRI systems may incorrectly resume scans after power loss when tissue heating rates exceed safe limits, risking excessive heating if patients contact conductive materials or have implants.

    Product
    SIGNA Creator, SIGNA Explorer, NMRI systems
    Category
    Distribution
    0 states
GE Healthcare (China) Co., Ltd. recalled 5 products on 2025-03-05 · The Recall Desk