FDA (Devices) recall action 96200

Medtronic MiniMed, Inc. recalled 12 products on 2025-03-12

The FDA (Devices) publishes a recall like this one as 12 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 12 pages.

Products
12
Announced
2025-03-12
Critical
1
Units
912,119

Why these products were recalled

All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. During a delivery accuracy dry run testing …

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–12 of 12

  • HighFDA (Devices)··2025-03-12

    MiniMed Insulin Pump Models Recalled for Abnormal Delivery During Flight

    Medtronic is recalling MiniMed insulin pumps due to air pressure changes during airplane takeoff and landing that can cause abnormal insulin delivery. Overdoses risk severe hypoglycemia; underdoses risk severe hyperglycemia and diabetic ketoacidosis.

    Product
    MiniMed insulin pump, REF: 740G (MMT-1811, MMT-1812, MMT-1861, MMT-1862)
    Category
    Distribution
    53 states
  • HighFDA (Devices)··2025-03-12

    Medtronic Paradigm insulin pump recall for abnormal delivery during flight

    Medtronic MiniMed Paradigm insulin pumps may deliver incorrect insulin doses during airplane takeoff or landing due to air pressure changes. This can cause severe high or low blood sugar requiring immediate medical attention.

    Product
    Paradigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT-751
    Category
    Distribution
    53 states
  • HighFDA (Devices)··2025-03-12

    Medtronic MiniMed insulin pump recalled for air pressure-triggered delivery defect

    Medtronic MiniMed insulin pumps can deliver abnormal insulin doses during air pressure changes in airplane travel, risking severe hypoglycemia or hyperglycemia and serious complications.

    Product
    MiniMed insulin pump, REF: 670G (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)
    Category
    Distribution
    53 states
  • HighFDA (Devices)··2025-03-12

    MiniMed Insulin Pumps Recalled for Abnormal Delivery During Air Travel

    Medtronic recalled MiniMed insulin pumps due to abnormal delivery caused by air pressure changes during airplane flights. The defect can cause severe hypoglycemia or hyperglycemia, potentially leading to seizure, coma, or death.

    Product
    MiniMed insulin pump, REF: 620G (MMT-1750)
    Category
    Distribution
    53 states
  • HighFDA (Devices)··2025-03-12

    Medtronic MiniMed Insulin Pumps Recalled for Abnormal Delivery During Air Pressure Changes

    Medtronic is recalling MiniMed insulin pumps that malfunction during airplane takeoff and landing due to air pressure changes. The defect can cause abnormal insulin delivery, leading to severe hypoglycemia, hyperglycemia, seizure, coma, or death.

    Product
    MiniMed insulin pump, REF: 640G (MMT-1711, MMT-1712, MMT-1751, MMT-1752)
    Category
    Distribution
    53 states
  • HighFDA (Devices)··2025-03-12

    MiniMed Insulin Pump May Malfunction at High Altitude During Flight

    Medtronic MiniMed insulin pumps may deliver abnormal insulin doses during airplane takeoff or landing due to air pressure changes, potentially causing severe low or high blood sugar.

    Product
    MiniMed insulin pump, REF: 770G (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
    Category
    Distribution
    53 states
  • HighFDA (Devices)··2025-03-12

    Medtronic Paradigm insulin pumps may deliver incorrect doses during flight

    Medtronic MiniMed insulin pumps may deliver abnormally high or low insulin doses in response to air pressure changes during airplane takeoff and landing, potentially causing severe hypoglycemia or hyperglycemia.

    Product
    Paradigm REF: MMT-712 and MMT-715
    Category
    Distribution
    53 states
  • SevereFDA (Devices)··2025-03-12

    Insulin Pump Recalled for Abnormal Delivery During Flight Pressure Changes

    Medtronic MiniMed 720G insulin pumps may malfunction during airplane travel due to air pressure changes, causing abnormal insulin doses that can lead to severe low or high blood sugar emergencies.

    Product
    MiniMed insulin pump, REF: 720G (MMT-1809, MMT-1810, MMT-1859, MMT-1860, MMT-1867)
    Category
    Distribution
    53 states
  • HighFDA (Devices)··2025-03-12

    Medtronic MiniMed insulin pumps may deliver abnormal insulin doses during flight

    Medtronic MiniMed insulin pumps may deliver abnormal doses of insulin during airplane takeoff or landing due to air pressure changes. This can result in severe hypoglycemia or hyperglycemia, with potential for seizure, coma, or death.

    Product
    MiniMed insulin pump, REF: 780G (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
    Category
    Distribution
    53 states
  • HighFDA (Devices)··2025-03-12

    MiniMed Insulin Pumps Recalled for Abnormal Insulin Delivery During Flight

    MiniMed insulin pumps may deliver abnormal insulin doses during aircraft takeoff and landing due to air pressure changes, risking severe blood sugar fluctuations, seizures, and potentially fatal complications.

    Product
    MiniMed insulin pump, REF: 630G (MMT-1714, MMT-1715, MMT-1754, MMT-1755)
    Category
    Distribution
    53 states