Why these products were recalled

Hollister received reports of decreased skin barrier wear time which could can lead to tube migration

The agency’s own wording, quoted from the enforcement report.

1–4 of 4

  • HighFDA (Devices)··2025-02-26

    AnchorFast Guard Endotracheal Tube Fasteners Recalled Due to Tube Migration Risk

    Hollister is recalling AnchorFast Guard endotracheal tube fasteners because decreased skin barrier wear time could cause tube migration. The recall affects multiple lot numbers distributed in the US and Canada.

    Product
    AnchorFast Guard, Oral Endotracheal Tube Fastener, REF 9800,
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-02-26

    Endotracheal Tube Fastener Recalled for Potential Tube Migration Risk

    Hollister is recalling AnchorFast SlimFit oral endotracheal tube fasteners worldwide due to decreased skin barrier wear time that could lead to tube migration.

    Product
    AnchorFast SlimFit, Oral Endotracheal Tube Fastener, REF 9787,
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-02-26

    Endotracheal Tube Fastener Recalled for Decreased Skin Barrier Wear

    Hollister is recalling AnchorFast Oral Endotracheal Tube Fasteners due to reports of decreased skin barrier wear time, which could lead to tube migration. The recall affects approximately 366,324 units distributed in the US and Canada.

    Product
    AnchorFast, Oral Endotracheal Tube Fastener, REF 9799,
    Category
    Distribution
    0 states