The Recall Desk

FDA (Devices) recall action 96168

Micro Therapeutics, Inc. recalled 2 products on 2025-03-12

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2025-03-12
Critical
0
Units
26,065

Why these products were recalled

Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • SevereFDA (Devices)··2025-03-12

    Medtronic Pipeline Vantage Embolization Device Recall for Structural Deformation Risk

    Medtronic is recalling approximately 18,245 unused Pipeline Vantage embolization devices due to incomplete wall apposition and braid deformation that may lead to thrombosis and serious adverse events.

    Product
    Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-20, PED3-027-350-25, PED3-027-400-12, PED3-027-400-14, PED3-027-400-16, PED3-027-400-18, PED3-027-400-20, PED3-027-400-25, PED3-027-400-30…
    Category
    Distribution
    48 states
  • SevereFDA (Devices)··2025-03-12

    Medtronic Pipeline Vantage device recall: deformation may cause blood clots

    The Pipeline Vantage Embolization Device may malfunction, with the braid deforming or not positioning correctly, potentially causing blood clots. Approximately 7,820 units are affected worldwide.

    Product
    Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PED3-021-275-12, PED3-021-275-16, PED3-021-325-12, PED3-021-325-14, PED3-021-325-16, PED3-021-325-20…
    Category
    Distribution
    48 states