The Recall Desk

FDA (Devices) recall action 96155

Augustine Temperature Management, LLC recalled 3 products on 2025-02-26

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2025-02-26
Critical
0
Units
8

Why these products were recalled

There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • HighFDA (Devices)··2025-02-26

    Patient Warming Mattress Recalled for Cable Connection Defect

    Augustine Temperature Management recalls 3 Hot Dog patient warming mattresses (US530) due to cable connector deflection that may cause intermittent connection failure, potentially occurring after one year of use.

    Product
    Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 89 cm (35"), REF US530, to prevent or treat hypothermia
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2025-02-26

    Hot Dog Patient Warming Mattress Connector Cable Deflection Recall

    Augustine Temperature Management is recalling 4 units of the Hot Dog Patient Warming Mattress due to connector cable deflection that causes intermittent connection failures after approximately one year of use.

    Product
    Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2025-02-26

    Patient Warming Mattress Connector Cable Intermittent Connection Defect

    Augustine Temperature Management is recalling Hot Dog Patient Warming Mattresses due to connector cable defects that may cause intermittent electrical connection after extended use, potentially affecting the device's heating function.

    Product
    Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 127 cm (50"), REF US502, to prevent or treat hypothermia
    Category
    Distribution
    3 states