MEDLINE INDUSTRIES, LP - Northfield recalled 3 products on 2025-02-05

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2025-02-05
Critical
0
Units
2,449

Why these products were recalled

Certain Medline kits were manufactured and distributed with Procedure Masks with SO SOFT Earloops, and SOFT TOUCH II Surgical Masks which were subsequently recalled for not meeting the bioburden testing requirements of EN 14683.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • HighFDA (Devices)··2025-02-05

    Medline procedure kits recalled for surgical masks failing sterilization tests

    Medline SHORT TERM INFUSAPORT ACCESS KIT with reorder number DYNDC3188 is recalled because included surgical masks failed bioburden testing, potentially compromising sterility during procedures.

    Product
    Medline procedure kits labeled as: 1) SHORT TERM INFUSAPORT ACCESS KIT, Reorder number DYNDC3188; surgery tray
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-02-05

    Medline Procedure Kits Recalled for Surgical Mask Bioburden Testing Failure

    Medline has recalled 749 procedure kits containing surgical masks that failed to meet bioburden testing standards under EN 14683. The affected kits were distributed nationwide.

    Product
    Medline procedure kits labeled as: 1) Port Access Kit, Reorder number CVI5130; 2) VANTEX 7FR 3L 20CM CVC STOCK+ BUNDLE US. Reorder number STCVC01USGV; general surgical tray
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2025-02-05

    Medline procedure kits recalled for surgical mask bioburden contamination risk

    Medline procedure kits containing surgical masks are recalled nationwide because included masks failed to meet bioburden testing standards, creating potential infection risk during medical procedures.

    Product
    Medline procedure kits labeled as: 1) CENTRAL LINE DRESSING KIT, Reorder number DYNDC3189; dressing change tray
    Category
    Distribution
    0 states