FDA (Devices) recall action 96031

Ortho-Clinical Diagnostics, INc. recalled 3 products on 2025-01-29

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2025-01-29
Critical
0
Units

Why these products were recalled

The VITROS 3600 Immunodiagnostic System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired products, which would increase the risk of no results or erroneous results from the instrument.

The VITROS XT 7600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired products, which would increase the risk of no results or erroneous results from the instrument.

The VITROS 5600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired products, which would increase the risk of no results or erroneous results from the instrument.

The agency’s own wording, quoted from the enforcement report.

1–3 of 3

  • HighFDA (Devices)··2025-01-29

    VITROS 3600 Diagnostic System May Fail to Alert Users to Expired Reagents

    The VITROS 3600 Immunodiagnostic System may not properly alert users when Signal Reagent Packs expire, risking use of expired products and potentially inaccurate test results. Affected systems worldwide should be checked for this software issue.

    Product
    Brand Name: VITROS Product Name: VITROS 3600 Immunodiagnostic System Model/Catalog Number: 6802783 (New), 6802914 (Certified/Refurbished), 6904000 (China) Software Version: version 3.8.0 or higher Product Description: In vitro diagnostic chemistry analyzer. FDA 3 letter pr
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-01-29

    Diagnostic analyzer software may fail to alert users to expired reagents

    VITROS XT 7600 Integrated System software may not warn users when Signal Reagent Packs expire, risking inaccurate or missing test results.

    Product
    Brand Name: VITROS Product Name: VITROS XT 7600 Integrated System Model/Catalog Number: 6844461 (New), 6272222 (Certified), 6904003 (China) Software Version: version 3.8.0 or higher Product Description: In vitro diagnostic chemistry analyzer. FDA 3 letter product code: JJE
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-01-29

    Medical Analyzer Software May Fail to Alert of Expired Reagent Packs

    The VITROS 5600 Integrated System's software (version 3.8.0 and higher) may fail to notify users when Signal Reagent Packs expire. This could result in use of expired reagents, causing no results or erroneous results from tests.

    Product
    Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: 6802413 (New), 6802415 (Certified/Refurbished), 6904001 (China) Software Version: version 3.8.0 or higher Product Description: In vitro diagnostic chemistry analyzer. FDA 3 letter product