The Recall Desk

FDA (Devices) recall action 95904

QUIDEL ORTHO recalled 2 products on 2025-01-22

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2025-01-22
Critical
0
Units
—

Why these products were recalled

Hemoglobin interference may occur at concentrations lower than the claimed 300 mg/dL, and may cause negatively biased digoxin results.

The agency’s own wording, quoted from the enforcement report.

1–2 of 2

  • HighFDA (Devices)··2025-01-22

    Digoxin Test Slides May Report Inaccurately Due to Hemoglobin Interference

    QUIDEL ORTHO is recalling VITROS DGXN test slides because hemoglobin can interfere with digoxin measurements at lower-than-expected levels, potentially causing inaccurate test results.

    Product
    Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides Model/Catalog Number: 8343386 Software Version: Not Applicable Product Description: The VITROS DGXN Slide is a multilayered, analytical element coated on a polyester support.…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-01-22

    VITROS DGXN Slides Recalled Due to Hemoglobin Interference in Digoxin Measurement

    VITROS DGXN Slides are recalled because hemoglobin can interfere with digoxin measurement, producing artificially low results. Approximately 1,541 units were distributed in the US and internationally.

    Product
    Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides Model/Catalog Number: 8343386 Software Version: Not Applicable Product Description: The VITROS DGXN Slide is a multilayered, analytical element coated on a polyester support.…
    Category
    Distribution
    0 states