MEDLINE INDUSTRIES, LP - Northfield recalled 18 products on 2025-01-08

The FDA (Devices) publishes a recall like this one as 18 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 18 pages.

Products
18
Announced
2025-01-08
Critical
0
Units
813,917

Why these products were recalled

Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–18 of 18

  • HighFDA (Devices)··2025-01-08

    Medical Device Kits Recalled for Non-Sterile Surgical Components

    Medline Industries is recalling See DocMan custom convenience kits that contain non-sterile Codman Surgical Patties and Strips with potential endotoxin contamination. Approximately 48,655 units were distributed worldwide.

    Product
    See DocMan
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-01-08

    Medline Surgical Procedure Kits Recalled for Non-Sterile Components

    Medline is recalling 943 custom surgical kits containing non-sterile Codman Surgical Patties and Strips that may be contaminated with endotoxin.

    Product
    Medline custom medical procedure kits labeled as: 1) FREE FLAP BASIC SET-UP PK, REF DYNJ0991320J; 2) MVFF/FIBU #54-RF, REF DYNJ21927R; 3) FUNCTIONAL SINUS ENDOSCOPY, REF DYNJ40406A; 4) DR GOLDSTEIN ACCESSORY PACK, REF DYNJ51818F; 5) ENDOSCOPIC SKULL BASE, REF…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-01-08

    Medline Deep Brain Stimulation Surgical Kits Recalled for Non-Sterile Contamination

    Medline has recalled 128 Deep Brain Stimulation surgical kits for containing non-sterile Surgical Patties and Strips with potential endotoxin contamination. Affected units were distributed worldwide including the US, UAE, and Canada.

    Product
    Medline custom medical procedure kits labeled as: DEEP BRAIN STIMULATION PACK-LF
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-01-08

    Medline Surgical Procedure Kits Recalled for Non-Sterile Components

    Medline is recalling custom surgical procedure kits containing non-sterile surgical components with potential endotoxin contamination. The recall affects 1,402 units distributed in the US, Canada, and UAE.

    Product
    Medline custom medical procedure kits labeled as: 1) VP SHUNT CDS, REF CDS983470D; 2) 1/2X1/2 NEURO SPONGE PACK-LF, REF DYNJ0713177; 3) BURR HOLE SHUNT PACK, REF DYNJ56819D; 4) BURR HOLE SHUNT PACK, REF DYNJ56819F; 5) BURR HOLE SHUNT PACK, REF DYNJ56819G; 6) RO VP…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-01-08

    Medline Custom Surgical Procedure Kits Recalled for Potential Endotoxin Contamination

    Medline Industries is recalling custom medical procedure kits due to non-sterile Codman Surgical Patties and Strips with potential endotoxin contamination. Approximately 1,072 units were distributed in the US, Canada, and UAE.

    Product
    Medline custom medical procedure kits labeled as: 1) ENDOSCOPIC SKULL BASE, REF DYNJ58648B; 2) ENDOSCOPIC SKULL BASE, REF DYNJ58648C; 3) ENT ENDO SINUS WILMINGTON PACK, REF DYNJ66447; 4) ENDOSCOPIC SKULL BASE, REF DYNJ58648D; 5) ENDOSCOPIC SKULL BASE, REF DYNJ58648J.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-01-08

    Medline Medical Procedure Kits Recalled for Non-Sterile Component Contamination

    Medline Industries is recalling 2,664 custom medical procedure kits containing non-sterile surgical components with potential endotoxin contamination. The affected kits were distributed in the US, UAE, and Canada.

    Product
    Medline custom medical procedure kits labeled as: 1) ENSEMBLE RACHIS-LF, REF DYNJ47568B; 2) UROLOGY ROBOTIC, REF DYNJ55555D; 3) KIT,SPINE LH, REF DYNJ906448A.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-01-08

    Medline custom surgical procedure kits recalled for non-sterile components and endotoxin contamination

    Medline custom procedure kits containing non-sterile Codman Surgical Patties & Strips are being recalled due to potential endotoxin contamination. The recall affects 74 units distributed in the US, UAE, and Canada.

    Product
    Medline custom medical procedure kits labeled as: 1) UMC DR. BAKER SUPPLEMENT, REF DYNJ69476; 2) DIEP FLAP, REF DYNJ910120
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-01-08

    Medline custom surgical procedure kits recalled for non-sterile components

    Medline is recalling 182 custom surgical procedure kits containing non-sterile Codman Surgical Patties and Strips with potential endotoxin contamination. The kits were distributed in the US, UAE, and Canada.

    Product
    Medline custom medical procedure kits labeled as: 1) CERVICAL PACK-LF, REF DYNJ0867145G; 2) CERVICAL SPINE PACK HUM-LF, REF DYNJ35184F; 3) ANTERIOR CERVICAL-CORPECTOMY, REF DYNJ904827.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-01-08

    Medline surgical procedure kits recalled for non-sterile supplies

    Medline is recalling custom medical procedure kits containing non-sterile surgical supplies with potential endotoxin contamination. The affected kits were distributed nationwide and internationally.

    Product
    Medline custom medical procedure kits labeled as: 1) CATH LAB PACK-LF, REF DYNJ0382559M; 2) CRANI, REF DYNJ56001B; 3) GENDER AFFIRMING SURGERY, REF DYNJ66252; 4) GENDER AFFIRMING SURGERY, REF DYNJ66252A; 5) MAJOR NEURO PACK-LF, REF DYNJ0578916X.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-01-08

    Medline Custom Surgical Kits Recalled for Non-Sterile Components with Endotoxin Risk

    Medline Industries is recalling 3,903 custom medical procedure kits due to non-sterile Codman Surgical Patties and Strips with potential for endotoxin contamination. The recall affects products distributed in the US, UAE, and Canada.

    Product
    Medline custom medical procedure kits labeled as: 1) MEM MAXILLOFACIAL PACK-LF, REF DYNJ0660040R; 2) FREE FLAP PACK-LF, REF DYNJ42833A; 3) FREE FLAP PACK-LF, REF DYNJ42833B; 4) TYMPANOPLASTY PACK, REF DYNJ50692C; 5) MAJOR CRANIOTOMY SAFETY, REF DYNJ54736B; 6) FREE FLAP…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-01-08

    Medline MICRODISECTOMY PACK recalled for endotoxin contamination

    Medline custom MICRODISECTOMY PACK kits containing non-sterile Codman Surgical Patties and Strips with potential endotoxin contamination are being recalled. Approximately 1,494 units were distributed in the US, UAE, and Canada.

    Product
    Medline custom medical procedure kits labeled as: 1) MICRODISECTOMY PACK, REF DYNJ80585A; 2) MICRODISECTOMY PACK, REF DYNJ80585B.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-01-08

    Medline Surgical Procedure Kits Recalled for Endotoxin Contamination Risk

    Medline Industries is recalling 2344 custom surgical kits with non-sterile components that may contain endotoxins. The kits were distributed in the US, UAE, and Canada.

    Product
    Medline custom medical procedure kits labeled as: 1) OPEN HEART CDS-LF, REFCDS840451M; 2) NEURO PACK, REFDYNJ41516M; 3) RFD- AAA PACK, REFDYNJ54242B; 4) OPEN HEART A&B, REFDYNJ901203C; 5) OPEN HEART A&B, REFDYNJ901203D; 6) ACDF-LF, REFDYNJ906681B; 7) ACDF-LF…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-01-08

    Medline Nasal Pack Procedure Kits Recalled for Potential Endotoxin Contamination

    Medline custom nasal pack procedure kits contain non-sterile surgical components with potential endotoxin contamination. Customers should discontinue use and contact Medline.

    Product
    Medline custom medical procedure kits labeled as: 1) NASAL PACK, REF DYNJ67789A.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-01-08

    Medline surgical procedure kits recalled for non-sterile components and endotoxin contamination

    Medline is recalling 474,300 custom surgical procedure kits because they contain non-sterile Codman Surgical Patties and Strips with potential endotoxin contamination. These components could pose infection risks during surgical procedures.

    Product
    Medline custom medical procedure kits labeled as: 1) CRANI, REF CDS780089V; 2) CRANIOFACIAL CDS, REF CDS780102K; 3) CRANIOFACIAL CDS, REF CDS780102L; 4) CRANIOFACIAL CDS, REF CDS780102M; 5) CRANIOTOMY CDS-LF, REF CDS780105K; 6) CRANIOTOMY CDS-LF, REF CDS780105L; 7)…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-01-08

    Medline surgical procedure kits recalled for potential endotoxin contamination

    Medline is recalling three custom surgical procedure kits due to non-sterile Codman Surgical Patties and Strips that pose a potential endotoxin contamination risk. The affected kits were distributed in the US, Canada, and the UAE.

    Product
    Medline custom medical procedure kits labeled as: 1) PANENDOSCOPY-LF, REF DYNJ36728C; 2) ENDOSCOPIC SINUS PACK, REF DYNJ40644A; 3) ENDO SINUS PACK, REF DYNJ66449.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-01-08

    Medline Surgical Procedure Kits Recalled Due to Sterility Concerns

    Medline is recalling custom surgical procedure kits that contain non-sterile surgical supplies with potential for endotoxin contamination. The kits were distributed worldwide, including the United States, United Arab Emirates, and Canada.

    Product
    Medline custom medical procedure kits labeled as: 1) NEURO BASIN PACK-LF, REF DYNJ39461J; 2) NEURO BASIN KIT, REF DYNJ80200B.
    Category
    Distribution
    0 states