FDA (Devices) recall action 95846

GE OEC Medical Systems, Inc recalled 4 products on 2024-12-25

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2024-12-25
Critical
0
Units
287

Why these products were recalled

Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–4 of 4

  • HighFDA (Devices)··2024-12-25

    GE OEC 9800 X-Ray System Recalled for Insufficient Tube Sealing

    GE OEC 9800 fluoroscopic X-ray systems are recalled due to insufficient sealing of X-ray tubes, which can result in oil leaks and loss of imaging capability. 25 systems distributed in the U.S. and internationally are affected.

    Product
    OEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The Series OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical application may include cholangiography, endoscop
    Category
    Distribution
    45 states
  • HighFDA (Devices)··2024-12-25

    Fluoroscopic X-Ray System Recalled for Insufficient X-Ray Tube Sealing

    GE OEC Elite mobile fluoroscopic X-ray systems are recalled due to insufficient sealing of X-ray tubes, which can cause oil leaks and loss of imaging capability. Approximately 100 systems have been affected.

    Product
    OEC Elite REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical
    Category
    Distribution
    45 states
  • HighFDA (Devices)··2024-12-25

    OEC 3D Mobile Fluoroscopic X-Ray System Tube Sealing Defect Recall

    GE OEC Medical Systems is recalling 84 OEC 3D mobile fluoroscopic X-ray systems due to manufacturing defects in X-ray tube sealing that can cause oil leaks and loss of imaging capability.

    Product
    OEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-Arm imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic Xray images and volumetric reconstructions durin
    Category
    Distribution
    45 states
  • HighFDA (Devices)··2024-12-25

    OEC 9900 Mobile Fluoroscopic X-Ray System Recall for Sealing Defect

    GE OEC Medical Systems is recalling 78 OEC 9900 Elite mobile fluoroscopy systems due to insufficient sealing of X-ray tubes. The defect can result in oil leaks and loss of fluoroscopic imaging capability.

    Product
    OEC 9900 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 9900 Elite mobile fluoroscopy system is designed to provide fluoroscopic and spot-film images of the patient during diagnostic surgical and interventional procedures. Examples of clinical application ma
    Category
    Distribution
    45 states