The Recall Desk

FDA (Devices) recall action 95780

Abiomed, Inc. recalled 2 products on 2025-01-22

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2025-01-22
Critical
1
Units
3,924

Why these products were recalled

Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary Artery (PA) Placement Signal & Pulmonary Artery Pulse Index (PAPi) metrics. Abiomed is reinforcing that there is a risk of potential interaction between the tip of guidewires, indwelling central venous lines or devices and inlet of the Impella pumps listed above during the insertion, manipulation, and removal of those devices. The interaction may result in optical sensor damage, t…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • SevereFDA (Devices)··2025-01-22

    Impella RP Heart Pump Recalled Due to Optical Sensor Damage Risk

    Abiomed is recalling 1,560 units of the Impella RP with SmartAssist pump due to risk of optical sensor damage from device interactions. Damaged sensors may cause temporary or permanent pump malfunction.

    Product
    Impella RP with SmartAssist; Product Number: 0046-0035;
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2025-01-22

    Impella RP Flex Devices Recalled Due to Optical Sensor Damage Risk

    Abiomed is recalling 2,364 units of the Impella RP Flex with SmartAssist due to risk of optical sensor damage from device interaction. Damage may cause temporary or permanent pump stop and loss of vital monitoring signals.

    Product
    Impella RP Flex with SmartAssist; Product Number: 1000323;
    Category
    Distribution
    0 states