The Recall Desk

FDA (Devices) recall action 95770

Trokamed GmbH recalled 4 products on 2025-01-22

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2025-01-22
Critical
0
Units
231

Why these products were recalled

If the inflow speed is too quick, it could result in reduced outflow through the shaft due to blood, tissue fragments and kidney stone fragments, and quickly build up of intrarenal pressure.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–4 of 4

  • SevereFDA (Devices)··2025-01-22

    Endoscopic Access Overtube Sheath Recalled Due to Pressure Buildup Risk

    Trokamed GmbH is recalling an endoscopic access overtube sheath due to risk of intrarenal pressure buildup when inflow speed is too fast. No injuries reported.

    Product
    Sheath, 18 Fr., 160 mm (REF: WA2PS18L), gastroenterology-urology endoscopic access overtube.
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2025-01-22

    Endoscopic Access Overtube Recalled for Risk of Obstruction and Pressure Buildup

    Trokamed is recalling a 20 Fr. endoscopic access overtube due to risk of obstruction and dangerous kidney pressure buildup during procedures. If inflow speed is too quick, rapid fluid flow can restrict outflow and cause intrarenal pressure to accumulate.

    Product
    Sheath, 20 Fr., 160 mm (REF: WA2PS20L), gastroenterology-urology endoscopic access overtube.
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2025-01-22

    Endoscopic Access Sheath Recalled Due to Risk of Intrarenal Pressure Buildup

    Trokamed GmbH is recalling an endoscopic access sheath (18 Fr., 130 mm) used in gastroenterology and urology. Rapid fluid inflow may block outflow and cause dangerous pressure buildup inside the kidney.

    Product
    Sheath, 18 Fr., 130 mm (REF: WA2PS18S), gastroenterology-urology endoscopic access overtube.
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2025-01-22

    Endoscopic Access Overtube May Cause Dangerous Pressure Buildup During Procedures

    Trokamed's endoscopic access overtube may cause rapid pressure buildup in the kidney if inflow speed is too fast. The device can become obstructed by tissue and blood fragments, restricting outflow.

    Product
    Sheath, 20 Fr., 130 mm (REF: WA2PS20S), gastroenterology-urology endoscopic access overtube.
    Category
    Distribution
    1 state