Critical
0
Units
277

Why these products were recalled

The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–6 of 6

  • HighFDA (Devices)··2024-12-25

    K-Systems T47 Warming Plate may exceed desired temperatures

    The K-Systems T47 Warming Plate has a temperature control defect where warming surfaces may exceed desired set temperatures. CooperSurgical is recalling 22 affected units distributed to fertility clinics.

    Product
    Brand Name: K-Systems Product Name: T47 Warming Plate Model/Catalog Number: K23054 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes
    Category
    Distribution
    14 states
  • HighFDA (Devices)··2024-12-25

    K-Systems G85 Mini Incubator May Exceed Desired Temperature

    CooperSurgical recalls 37 K-Systems G85 Mini Incubators because the device warming surfaces may exceed the desired set temperature. The unit alarms when this occurs, but affected serial numbers require verification.

    Product
    Brand Name: K-Systems Product Name: G85 Mini Incubator Model/Catalog Number: K22074 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes
    Category
    Distribution
    14 states
  • HighFDA (Devices)··2024-12-25

    K-Systems Incubator Device Recall: Temperature May Exceed Desired Setting

    CooperSurgical is recalling 34 K-Systems incubator devices used in assisted reproduction procedures. The warming surfaces may exceed the desired set temperature, though the device will alarm if this occurs.

    Product
    Brand Name: K-Systems Product Name: Multiple/unknown Model/Catalog Number: Multiple/unknown Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g.,
    Category
    Distribution
    14 states
  • HighFDA (Devices)··2024-12-25

    K-Systems G73 Dry Bath Incubator may exceed target temperature

    CooperSurgical is recalling K-Systems G73 Dry Bath Incubators because warming surfaces may exceed the desired set temperature. The device alarms when this occurs.

    Product
    Brand Name: K-Systems Product Name: G73 Dry Bath Incubator Model/Catalog Number: K22065 Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at c
    Category
    Distribution
    14 states
  • HighFDA (Devices)··2024-12-25

    K-Systems T47 Warming Plate May Exceed Set Temperatures

    CooperSurgical recalls K-Systems T47 Warming Plates because temperature may exceed the desired set value, potentially affecting biological materials. The device alarms appropriately when this occurs.

    Product
    Brand Name: K-Systems Product Name: T47 Warming Plate Model/Catalog Number: K23055 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes
    Category
    Distribution
    14 states
  • ModerateFDA (Devices)··2024-12-25

    K-Systems R65 embryo incubator temperature may exceed setpoint

    CooperSurgical's K-Systems R65 Trolley reproductive incubators may experience temperature excursions above desired setpoints, though devices will alarm when this occurs. The FDA issued a Class II recall of 43 units across eight U.S. states.

    Product
    Brand Name: K-Systems Product Name: R65 Trolley Model/Catalog Number: K24012 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and em
    Category