Why these products were recalled

Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • HighFDA (Devices)··2024-12-04

    Smiths Medical Tracheal Tube Exchange Guide Recalled for Inadequate Disinfection Risk

    Smiths Medical recalled its PORTEX Tracheal Tube Exchange Guide due to potential fluid ingress during reprocessing that may compromise disinfection standards and pose a contamination risk.

    Product
    smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-12-04

    Tracheal Tube Exchange Guide Recalled for Inadequate Disinfection Risk

    Smiths Medical is recalling PORTEX Tracheal Tube Exchange Guides due to potential fluid ingress during reprocessing and inadequate disinfection per applicable standards.

    Product
    smiths medical PORTEX Tracheal Tube Exchange Guide, 10Ch x 70cm, REF 14-504-76
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-12-04

    Smiths Medical Tracheal Tube Introducer Recall Due to Inadequate Disinfection

    Smiths Medical is recalling reusable tracheal tube introducers due to potential fluid ingress during reprocessing and concerns about the adequacy of recommended disinfectants.