The Recall Desk

FDA (Devices) recall action 95486

Philips Respironics, Inc. recalled 7 products on 2024-11-20

The FDA (Devices) publishes a recall like this one as 7 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 7 pages.

Products
7
Announced
2024-11-20
Critical
0
Units
154,564

Why these products were recalled

Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–7 of 7

  • SevereFDA (Devices)··2024-11-20

    Trilogy Evo Universal Ventilator Recalled Due to Sensor Malfunction Risk

    Trilogy Evo Universal Ventilators (Model DS2000X11B) are recalled because in-line nebulizer configuration in certain locations could cause aerosol accumulation on the internal flow sensor, potentially resulting in inaccurate measurements.

    Product
    Trilogy Evo Universal Ventilator. Model Number DS2000X11B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.
    Category
    Distribution
    12 states
  • SevereFDA (Devices)··2024-11-20

    Aeris EVO Ventilator May Develop Flow Sensor Fouling from Aerosols

    Philips Respironics Aeris EVO ventilators may accumulate aerosol deposits on the internal flow sensor, potentially causing inaccurate flow measurements.

    Product
    Aeris EVO. Model Number: VT2110X24B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.
    Category
    Distribution
    12 states
  • SevereFDA (Devices)··2024-11-20

    Philips LifeVent EVO2 ventilators recalled for potential sensor accuracy failure

    Philips Respironics is recalling LifeVent EVO2 ventilators due to potential aerosol deposit accumulation on the internal flow sensor that could cause inaccurate measurements during ventilation.

    Product
    LifeVent EVO2. Model Number: SP2100X26B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.
    Category
    Distribution
    12 states
  • SevereFDA (Devices)··2024-11-20

    Trilogy EV300 ventilator: potential flow sensor malfunction from aerosol accumulation

    Philips Respironics is recalling 34,146 Trilogy EV300 ventilators due to potential aerosol accumulation on the internal flow sensor, which could cause inaccurate ventilation measurements.

    Product
    Trilogy EV300. Model Numbers: DS2200X11B, IN2200X15B, BL2200X15B, BR2200X18B, CA2200X12B, DE2200X13B, GB2200X15B, ND2200X15B, ES2200X15B, FR2200X14B, FX2200X15B, EU2200X15B, IA2200X15B, IT2200X21B, KR2200X15B, TR2200X15B, EE2200X15B. Intended for invasive and non-invasive venti
    Category
    Distribution
    12 states
  • SevereFDA (Devices)··2024-11-20

    Ventilator Model LD2110X23B Recalled for Potential Sensor Measurement Failure

    Philips Respironics is recalling 1,322 Garbin Evo ventilators (Model LD2110X23B) worldwide due to potential aerosol accumulation that could cause inaccurate flow sensor readings, posing a risk to patient care.

    Product
    Garbin Evo. Model Number: LD2110X23B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.
    Category
    Distribution
    12 states
  • SevereFDA (Devices)··2024-11-20

    Trilogy Evo O2 Ventilators Recalled for Potential Flow Sensor Malfunction

    Philips Respironics recalled 24,249 Trilogy Evo O2 ventilators because in-line nebulizer configurations may allow aerosol deposits to accumulate on the device's flow sensor, potentially causing inaccurate flow measurements.

    Product
    Trilogy Evo O2. Model Numbers: FP2100X10, DS2100X11B, IN2100X15B, DE2100X13B, IA2100X15B, EU2100X15B, LA2100X15B, EE2100X15B, FX2100X15B, ES2100X15B, ND2100X15B, FR2100X14B, IT2100X21B, JP2100X16B, IN2100X19, EU2100X19. Intended for invasive and non-invasive ventilation of pedi
    Category
    Distribution
    12 states
  • SevereFDA (Devices)··2024-11-20

    Philips Trilogy Evo ventilators recall: flow sensor fouling risk

    Philips Respironics is recalling Trilogy Evo ventilators due to potential aerosol deposits on the internal flow sensor, which could cause inaccurate flow measurements during ventilation.

    Product
    Trilogy Evo. Model Numbers: DS2110X11B, UDS2110X11B, RDS2110X11B, CA2110X12B, JP2110X16B, IN2110X15B, DE2110X13B, AU2110X15B, BL2110X15B, EU2110X15B, BR2110X18B, LA2110X15B, EE2110X15B, FR2110X14B, GB2110X15B, ND2110X15B, ES2110X15B, IA2110X15B, IT2110X21B, KR2110X15B, TR2110X15B
    Category
    Distribution
    12 states