FDA (Devices) recall action 95377

Luminex Molecular Diagnostics Inc recalled 2 products on 2024-10-16

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2024-10-16
Critical
0
Units
3,988

Why these products were recalled

Respiratory pathogen panel may report an elevated rate (7.6% predicted probability of occurrence) of non-specific hMPV target (human Metapneumovirus) results, risk identified predominantly in clinical samples that present positivity for another target, leading to an erroneous co-infection result, can potentially result in hMPV false positives.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2024-10-16

    Respiratory Pathogen Panel Recalled Due to False Positive Results

    The NxTAG Respiratory Pathogen Panel is recalled due to false positive results for human metapneumovirus in about 7.6% of samples, especially those positive for other respiratory pathogens.

    Product
    NxTAG Respiratory Pathogen Panel, REF: I051C0447
    Category
    Distribution
    19 states